Page | 35
Commercial in confidence. User Manual Revision CTK-SM-REV8-060720
6. SERVICING INSTRUCTIONS
- Continued
6.21 Roving Handset - Replacement
(Refer Figure 4.)
Removal
a) Raise the mattress platform to maximum height and remove seat cover (figure 7
item 41). Unplug the bed from mains power supply socket.
b) Release junction box (8) cable retainer by inserting a small flat screwdriver into the
slots (A) on the front of the junction box and pry off tabs (B)
– refer figure 18
c) Pull off the face plate and put it aside.
d) Unplug the handset (19 & 20) control cables from the junction box.
e) Cut and remove cable ties securing the handset cable to the bed frame.
Installation
a) Route the handset cable (19 & 20) from the required side of the bed to the junction
box (8).
b) Plug in the handset cable back into the original junction box port.
c) Install the cable retainer face plate.
d) Secure the handset cable to the bed frame using cable ties.
e) Reinstall seat cover (figure 7 item 41).
6.22 Attendant Control Panel (ACP) - Replacement
(Refer Figure 4.)
Removal
a) Raise the mattress platform to the maximum height and unplug the bed from mains
power supply wall socket.
b) Insert a small flat screwdriver into the slot (A) on the outside of the cable inlet socket
of the attendant control panel (21) to release the cable retaining clip. Whilst pushing
down on the tab with the screwdriver, pull out the cable from the attendant control
panel socket.
c) Remove two self-tapping screws securing the control panel to the bed frame.
Installation
a) Plug the cable into the attendant
control panel (21) socket. Make sure
that the retaining clip clicks into place
and that the cable cannot be
accidentally pulled out.
b) Secure the control panel to the bed
frame by reinstalling the two self-
tapping screws.