Page | 38
Commercial in confidence. User Manual Revision CTK-SM-REV8-060720
6. SERVICING INSTRUCTIONS
- Continued
Installation
a) Apply grease to both pivot points prior to assembly.
b) Position actuator, ensuring correct orientation and attach to bed frame using
existing bolts and new Nyloc nuts (figure 7 items 27 & 39, figure 8 items 27 & 30).
Be sure not to over tighten pivot bolts allowing actuator to pivot freely.
c) Note: Actuator Safety spline feature allows for ram manual adjustment (by rotating
the ram) if required during this procedure.
d) Lightly lubricate actuator cable plug O ring (6) with petroleum jelly and install,
securing it in place with retaining clip.
e) Insert CPR release cable outer into actuator cable recess followed by inner cable
ball end and slide cover in position.
f) Install seat cover in position (figure 7 item 41).
6.27 Actuator Cable (common) - Replacement
(Refer figure 4)
Removal
a) Raise the bed to maximum height and disconnect bed from mains power supply.
b) For mattress platform sections actuator/s, lower the sections to a flat position and
remove ABS top cover/s to improve visibility and accessibility.
c) Remove actuator cable retaining clip from actuator housing port and disconnect
cable
– refer to section 6.1.2 figure 3.
d) Remove control box (1) from bed frame
– refer to subsection 18 - control box
replacement procedure.
e) Open control box lid using a small flat screwdriver and unplug cable to be replaced
(figure 4 item 6).
Installation
a) Lightly lubricate actuator cable (6) plug O ring with petroleum jelly and plug it into
the control box port (1). Clip the cable into the cable guide slot on the control box
housing and close the lid. Check that cable is properly secured.
b) Re-attach the control box to the mounting bracket on the bed frame
– refer to
subsection 18 control box replacement procedure.
c) Route cable from the control box along the bed frame to the actuator.
d) Lightly lubricate cable plug O ring with petroleum jelly and plug it into the actuator
port. Push down on the retaining clip to secure cable in position. Check that cable is
properly secured.
e) Apply cable ties to secure cable to the bed frame.
f) Reinstall mattress platform section cover/s.