How to use the Maxi 500
7
Adjusting Legs Spreading
Fig. 3
Close Legs
Open Legs
The legs’ opening width
can be adjusted by using
the two bottom buttons on
the hand control.
Brakes
)RRW RSHUDWHG EUDNHV DUH ¿WWHG RQ ERWK UHDU
castors.
To apply brakes
, step on the back portion of
pad.
To release brakes
, push the top portion of the
pad forward.
Fig. 4
Emergency Stop
The operator can shut off the power at any time
by pressing the red emergency button on the
control panel or by pressing the red button on
the battery while pulling it backward to remove
it.
Reset the emergency stop function by pressing
on the green power button or by replacing the
battery.
First-time users should practice the emergency
stop manoeuvre before operating the lift with a
patient in it.
Fig. 5
Fig
F
F
F
. 5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
Emergency
Stop
Button
Battery
release
button
Pull
Boom Control
The electrical actuator can be operated in both
directions. It allows the operator to raise or
lower the patient without any physical effort.
7KHERRPUDLVLQJORZHULQJDFWLRQLVFRQWUROOHG
E\ WKH ³8S´³'RZQ´ EXWWRQV ORFDWHG RQ WKH
control box and on the hand control.
If two buttons are pressed simultaneously, the
¿UVWIXQFWLRQEXWWRQSUHVVHGRYHUULGHVWKHRWKHU
function until it is released.
Fig. 6
To Raise the Patient
WARNING
: Always make sure that the
spreader bar is above the patient before
lifting.
Failure to follow this instruction may cause
the patient to swing resulting in injury.
7KH³83´EXWWRQVDUHXVHGWRUDLVHWKHERRP
The boom keeps rising as long as the button is
pressed or until it reaches its upper travel limit.
To Lower the Patient
The “Down” buttons are used to lower the boom.
The boom keeps lowering as long as the button
is pressed or until it reaches its lower travel
limit.
How to use the
Maxi 500
Raise Patient
Lower Patient
Locked
8QORFNHG