Electromagnetic Compatibility
29
Electromagnetic Compliance
The
Maxi 500
has been tested for compliance with current regulatory standards regarding its
capacity to block EMI (electromagnetic interference) from external sources.
Nonetheless, some procedures can help reduce electromagnetic interferences:
•
(QVXUHWKDWRWKHUGHYLFHVLQSDWLHQWPRQLWRULQJDQGRUOLIHVXSSRUWDUHDVFRPSO\WR
accepted emissions standards.
•
0D[LPL]HWKHGLVWDQFHEHWZHHQHOHFWURPHGLFDOLJKSRZHUHGGHYLFHVPD\
SURGXFH(0,WKDWFDQDIIHFWWKHÀRRUOLIW
For more information on how to manage the unit’s RF electromagnetic environment, please
consult the AMI TIR 18-1997 - Guidance on Electromagnetic Compatibility of Medical Devices for
&OLQLFDO%LRPHGLFDO(QJLQHHUV
WARNING
8VHRIDFFHVVRULHVFDEOHVDQGVSDUHSDUWVRWKHUWKDQWKRVHVSHFL¿HGRUSURYLGHG
by Arjo could result in increased electromagnetic emissions or decreased electromagnetic
immunity of this equipment and result in improper operation.
WARNING
8VH RI WKLV HTXLSPHQW DGMDFHQW WR RU VWDFNHG ZLWK RWKHU HTXLSPHQW VKRXOG EH
avoided because it could result in improper operation. If such use is necessary, this equipment
and the other equipment should be observed to verify that they are operating normally.
WARNING
: The equipment may cause radio interference or may disrupt the operation of
nearby equipment. It may be necessary to take action, such as reorienting, relocating the
equipment or shielding the location.
Electromagnetic Emissions
Guidance and Manufacturer’s Declaration -
Electromagnetic Emissions - For all Equipment and Systems
The
Maxi 500
is intended for use in the electromagnetic environment indicated below. The
FXVWRPHURUWKHXVHURIWKH0$;,VKRXOGDVVXUHWKDWLWLVXVHGLQVXFKDQHQYLURQPHQW
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The
Maxi 500
uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The
Maxi 500
is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.