MRM Carendo
Service Procedures
19
15.Remove the blue button, two screws inside
and three screws outside.
16.Disconnect the motor connections J8, J9 on
the PCBA.
17.Remove the cover.
18.Loosen the three bolts for the worm gear unit.
To replace only the motor:
19.Loosen the two bolts and pull off the motor
from the splines.
Reassembling:
1. When reassembling notice the position of the
cable.
2. Use Loctite according to page 24 and correct
torques according to page 30.
3. Be careful about the cable protection.
15.
A
B
D
D
C
A= Longest screw
B= Second longest screw
C= Shortest screw
D= Equal length of screws
15
17-18
19