MRM Carendo
Trouble shooting
31
Trouble shooting
Part Problem
Cause
Action
Service
procedure
Castors
Bad rolling
Dirt/wear
Replace castors
10
Bad swivel-
ling
Dirt/wear
Replace castors
10
Seat Seat
will
not recline
Fuse bad
Check fuse F1
11
Contacts in wrong
position
Check contacts on PCBA
PCBA J7,
J9, J10, J11
Bad connection
Check hand set and cables
PCBA J:1
Motor fault
Check worm gear unit;
motor
13
Microswitch fault
Check microswitch lying
PCBA
J11:1,2
Electronic fault
Check PCBA
11
Seat will
not rise
Fuse bad
Check fuse F1
11
Contacts in wrong
position
Check contacts on PCBA
J7, J9, J10,
J11
Bad connection
Check hand set and cables
PCBA J:1
Motor fault
Check worm gear unit;
motor
13
Microswitch fault
Check microswitch sitting
PCBA
J10:1,2
Electronic fault
Check PCBA
11
Seat will
recline/rise
but makes
noises
Bad attachment,
wear
Check upper tie rod
4
Bad lubrication/
attachment, wear
Check/replace bearings
seat
5
Bad lubrication/
attachment, wear
Check/replace worm gear
unit
13
NOTE
Before any further actions are taken: Check that the battery is OK and fully charged.
(SP2)