MRM Carendo
Service Procedures
23
SP 15. Replace emergency
buttons
1. Proceed according to page 17; item 1-7.
2. Study a button with cabling to find out the
best way to separate the button from the
connector using a small screwdriver and
lifting the latch at the same time rotating the
connector.
3. When the cable with connector is loose,
unscrew the locking ring
(a)
for the button
turning it counter-clockwise
Reassembling
4. When reassembling note that the peg in the
button and in the connector shall be placed
crosswise.
5. Inside the cover a rotation stop
(b)
has been
added 2013 for the emergency stop button.
Add if missing. Teeth against cover
6. Proceed according to page 17; item 11-14
NOTE
If you are unable to remove the button according
to this procedure, proceed according to page 18
.
2
a
3
b