11 Resistance symmetry (RSym)
ARGUS Copper Box
59
Resistance symmetry measurement settings
ARGUS in the status display.
The measurement is not yet started.
Hotkeys/settings of resistance symmetry
measurement s. page 20 and in the sheet
under.
Under certain circumstances, the resistance symmetry measurement can abort.
Possible causes include the presence of an external voltage, absence of a line or
of a loop.
When you select the susceptibility
"insecure (fast)", the red warning
"ATTENTION: susceptible"
appears on the display.
If a wire type has been selected,
this is shown in the display (in this
example wire type 1).
<Menu>
Opens the Copper Box
<Start>
Starts the resistance
symmetry measurement.
Settings
Description
Wire types:
See page 30 for an explanation of wire types.
Susceptibility:
Selects the susceptibility with which ARGUS performs resistance symmetry
measurement.
Robust for external
voltage:
Insecure (fast):
Default:
The measurement is reliable in the event of external voltages up to
30 V.
Using another measuring method, the measurement is extremely fast
but much more susceptible to interference.
Robust for external voltages