![ARDETEM PECA 30 Configuration Manual Download Page 39](http://html1.mh-extra.com/html/ardetem/peca-30/peca-30_configuration-manual_2966112039.webp)
PECA 30 / 30 E : Accessible functions during the measurement
39
ARDETEM
CURRENT CUT-OFF CALIBRATION
If the analyser displays a current value while the installation doesn’t flow, the analyser can be current zero
reset as bellow :
CAUTION :
the installation doesn’t flow
Press simultaneously the keys :
and
The analyser displays
during 2s
Back to the measurement with currents to 0.
In case of mistake, to go back to the default calibration, proceed the same :
Press simultaneously the keys :
and
The analyser displays
during 2s
Back to the measurement
Offset minimum by calibration : 70 mA for a 5A secondary CT
10 mA for a 1A secondary CT