PECA30 / 30
E : general presentation
16
ARDETEM
•
TIME METER DATE AND TIME : NOT AVAILABLE ON PECA 30 / 30 E
We have in option a time-meter function, which is supplied with the measurement data storage option.
PATTERN EXAMPLE : the 9th may 1997 it is 9h45min
Display configured with 3 numbers after the decimal point : Year = 1997.0 Day, Month = 9.050
Hour = 9.450
A display without decimal point for the year and 2 decimals for the day and the hour will give a readable
display : Year = 1997 Day, Month = 9.05 Hour = 9.45
•
MAXIMUM OF THE CURRENTS :
The analyser displays the maximum of the currents on the last N minutes (N is programmable
ex :10 minutes) (average value calculated on the last N minutes).
The duration N is common with the integration duration of the mean powers.
CAPTURE FUNCTIONS
Nr1 - 4 DIGITS NUMBER WITH SIGNE AND DECIMAL POINT :
4
th
digit + sign, from –9 to +9
3
rd
digit from 0 to 9
2
nd
digit from 0 to 9
1
st
digit from 0 to 9
decimal point