OPERATING INSTRUCTIONS
ORIX 70
The external “
ready
” light, if connected, lights on, showing the presetting of the equipment to take
the x-ray.
The “
ready
” light is not furnished while are already preset on the power cards the electric links.
TECHNICAL CONTROL DURING INSTALLATION
Check that power supply is 230 VAC +/- 10%
Check that all the functions relative to the setting of technical data are operating correctly:
a)
generator setting
b)
film sensitivity setting
c)
patient setting
d)
tooth setting
e)
“ENTER” setting
f)
Pause time
CAUTION!
This operation must be done before connecting the x-ray equipment.
To test it is necessary to use a sufficient power output supply
E.g. 300 W – 230 VAC halogen lamp.
Only if this test is positive the x-ray equipment can be connected according to the foreseen safety
rules.
For connection, refer to
Diagram C2