Proper usage of the Refrigerator
During the initial startup and continuous usage of the refrigerator, the following procedures should be
followed.
1. Power the refrigerator in a dedicated power socket. The voltage requirement is 220 VAC and 50 Hz.
When started, do not place any products in the refrigerator.
2. When the unit initially starts up, the buzzer alarm may sound. This is normal. You can push the
buzzer alarm button to cancel the sound. The buzzer alarm continues to work until the upper bottle
temperature reaches the range of 2° C to 6° C.
3. Make sure that both sensor bottles have glycerin solution of 10% concentration.
Please refer to item 4 in the Operation section for details.
4. This refrigerator has been factory preset to operate at a range of 4° C
. It is not necessary
to
adjust the temperature set point.
5. It usually takes several hours for the unit to reach the preset temperature. Once the unit’s
temperature is stable, check the sensor bottle temperature to see if it matches the set point.
6. Turn on the light for the interior. Inspect the light to make sure it functions properly.
7. Once a thorough inspection of the unit is completed, products can be loaded into the refrigerator.
Beware to not load too many warm products into the unit at one time.
Remote alarm contact
The terminals for the remote alarm contact are located on the back of the unit cover panel. The
contacts are designed for 30 VDC and 2A load. The remote alarm contacts work in synchronization
with the buzzer alarm on the refrigerator. Therefore, the remote alarm can be terminated by pushing the
buzzer alarm button. However, in case of a power outage, the remote alarm contacts cannot be
controlled by the buzzer alarm.
Remote alarm condition: Open for Normal, Close for Abnormal
+/- 2° C
23
Summary of Contents for BBR100
Page 1: ...OPERATING INSTRUCTIONS BLOOD BANK REFRIGERATOR 100 300 500 700 1400...
Page 10: ...Before you start the refrigirator and freezer you have to connect the battery backup 10...
Page 30: ...BBR Units Block Diagram of Main PCB board 30...
Page 31: ...Wiring diagram for BBR 500 700 ST 31...
Page 32: ...Wiring diagram for BBR 1400 ST 32...
Page 33: ...Notes 33...
Page 34: ...Notes 34...