Screen
Main menu
At alarm
Temp. Last 24H max/min
Settings level l password
Settings level ll password ll
Hysteresis 13)
Range
On/off
Calibration 17)
Automatic defrost 18)
Change passwords
Service menu
Reset all settings to fabric default 23)
Max temp limit 14)
Min temp limit 15)
Password protected yes/no 16)
Yes. Password
No
Yes
No
Defrost cyclus
Start manual defrost
Change password l 19)
Change password ll 20)
Compressor hours 21)
Fan hours 22)
Description control panel for engineer
13 Hysteresis
The difference between the set point and the temperature that
causes the fan and the compressor to restart.
14 Max temp limit
Max. set point allowed to be set on level I
15 Min temp limit
Min. set point allowed to be set on level II
16 Password protected Yes/no.
Password for start/stop the freeer Yes or No?
17 Calibration
The temperature sensor of the unit can be calibrated.
This value represents the /- from the value in the display.
18 Automatic defrost
Yes: Defrost cyclus
Adjust how many times the unit must automatic defrost.
No: manual defrost
The unit will only defrost when manual defrost is being
activated.
19 Change password l
Changes the password to level 1
20 Change password ll
Changes the password to level 2
21 Compressor hours
Compressor running time?
22 Fan hours
Fan running time
23 Reset all settings to fabric default
All settings will be deleted and turn back to factory settings.
16
Summary of Contents for BBR100
Page 1: ...OPERATING INSTRUCTIONS BLOOD BANK REFRIGERATOR 100 300 500 700 1400...
Page 10: ...Before you start the refrigirator and freezer you have to connect the battery backup 10...
Page 30: ...BBR Units Block Diagram of Main PCB board 30...
Page 31: ...Wiring diagram for BBR 500 700 ST 31...
Page 32: ...Wiring diagram for BBR 1400 ST 32...
Page 33: ...Notes 33...
Page 34: ...Notes 34...