Installation
1. Remove packaging materials, packaging bags and straps.
Open the refrigerator door to let it air out. If there are dirt marks on the outside of the unit, use a
soft sponge with a light detergent solution to wipe the unit clean. Then, rinse the unit with a clean,
slightly damp towel.
2. Adjust the height of the unit.
Turn the leveling leg screw counterclockwise to level the front end of the unit. The front casters
should be off the ground.
3. Preparation of sensor bottle.
Before using the refrigerator, inspect the sensor bottle located at the back of the drawers and
shelves to see if the liquid level is up to the 200 mark inside the bottle. If the bottle has no fluid or
the fluid level is lower than the 200 mark, use the following steps to fill the bottles at the upper
corner and the lower corner with solution of 10% glycerin concentration (or other equivalent liquid
such as glycerin solution).
(1) Remove the top drawer, bottom drawer, shelves and baskets.
(2) Unscrew the fastener that positions the bottle, remove the bottle.
(3) Unscrew the bottle’s cap that is used for the sensor mounting.
(4) Pour a solution of 10% glycerin concentration into the bottle(s) to the 200 mark line.
(5) Mount the bottle back to the refrigerator liner wall.
(6) Install the bottle back to its original position. Place the temperature sensor back inside the
temperature bottle.
14
Summary of Contents for BBR100
Page 1: ...OPERATING INSTRUCTIONS BLOOD BANK REFRIGERATOR 100 300 500 700 1400...
Page 10: ...Before you start the refrigirator and freezer you have to connect the battery backup 10...
Page 30: ...BBR Units Block Diagram of Main PCB board 30...
Page 31: ...Wiring diagram for BBR 500 700 ST 31...
Page 32: ...Wiring diagram for BBR 1400 ST 32...
Page 33: ...Notes 33...
Page 34: ...Notes 34...
Page 35: ...Notes 35...