Overview of messages, warnings and error
messages
Select surgeon!
Asks you to select the correct surgeon number
Pressure too low
The pressure at the compressed air connection is too low.
Venturi and VIT PN cannot be operated properly. Check
mains pressure and replace venturi nozzle filter if necessary
(Service Manual)
Connect handpiece!
The phaco handpiece is not connected, or not connected
properly.
Defective handpiece!
The handpiece cannot be operated.
See Section 8.3.1
Test handpiece!
The handpiece has a low output.
See chapter 8.3.1.
Insert cassette!
Insert a cassette, or close the cassette door properly.
Empty cassette!
The cassette is completely full. Empty immediately.
No pedal!
The pedal is not connected or not properly connected.
PREOP discontinued!
PREOP was interrupted by pressing pedal with the heel.
Phaco test active
The phaco test is under way and nothing else can be
operated.
Phaco test OK
The phaco test has been completed successfully.
Please repeat phaco test!
The test was interrupted and must be repeated.
Phaco Test discontiued!
The test was interrupted by pressing pedal with the heel.
System ready!
The self-test has been completed successfully.
Check tip!
The tip is not properly attached.
See Section 8.3.1
Call Service,
Error 1
Contact service (Internal voltages defective)
The unit cannot be used.
Adjust Unit,
Error 2
Contact service. (Pressure measurement incorrect)
The unit cannot be used.
Adjust Pedal
Error 3
Contact service (The pedal is not properly set)
The unit cannot be used.
Graphic Display
Error 4
Call service (ParaProg display is defective)
Program Failure
Error 8
Contact service. The unit cannot be used.
NV Ram failure
Error 9
Contact service. The unit cannot be used.
Pump Failure
Error 10
Contact service (Pump doesn't work).
The unit cannot be used.
Overload
Error 11
Loading too high. Wait until the respective switching circuit
has cooled down: the message will disappear.
Irr Motor
Error 12
Infusion rod has been knocked.
Wait until the message disappears, then lower the rod.
Update Vitrex
Error 13
BASIC has a newer SW then VITREX
Call service
Force Sensor
Error 15
Forcesensor defective.
Call service
Adjust Venturi
Error 16
Impossible venturi sensor calibration values.
Call service.
Adjust Force Sens
Error 17
Impossible peristaltic sensor calibration values.
Call service.
Check Venturi
Error 18
Venturi System: Vacuum deviation to high.
Call service
Errors 5, 6 and 7 do not appear for this unit