Overview of messages, warnings and error
messages

Select surgeon!
Asks you to select the correct surgeon number

Pressure too low
The pressure at the compressed air connection is too low.
Venturi and VIT PN cannot be operated properly. Check
mains pressure and replace venturi nozzle filter if necessary
(Service Manual)

Connect handpiece!
The phaco handpiece is not connected, or not connected
properly.

Defective handpiece!
The handpiece cannot be operated.
See Section 8.3.1

Test handpiece!
The handpiece has a low output.
See chapter 8.3.1.

Insert cassette!
Insert a cassette, or close the cassette door properly.

Empty cassette!
The cassette is completely full. Empty immediately.

No pedal!
The pedal is not connected or not properly connected.

PREOP discontinued!
PREOP was interrupted by pressing pedal with the heel.

Phaco test active
The phaco test is under way and nothing else can be
operated.

Phaco test OK
The phaco test has been completed successfully.

Please repeat phaco test!
The test was interrupted and must be repeated.

Phaco Test discontiued!
The test was interrupted by pressing pedal with the heel.

System ready!
The self-test has been completed successfully.

Check tip!
The tip is not properly attached.
See Section 8.3.1

Call Service,

Error 1

Contact service (Internal voltages defective)
The unit cannot be used.

Adjust Unit,

Error 2

Contact service. (Pressure measurement incorrect)
The unit cannot be used.

Adjust Pedal

Error 3

Contact service (The pedal is not properly set)
The unit cannot be used.

Graphic Display

Error 4

Call service (ParaProg display is defective)

Program Failure

Error 8

Contact service. The unit cannot be used.

NV Ram failure

Error 9

Contact service. The unit cannot be used.

Pump Failure

Error 10

Contact service (Pump doesn't work).
The unit cannot be used.

Overload

Error 11

Loading too high. Wait until the respective switching circuit
has cooled down: the message will disappear.

Irr Motor

Error 12

Infusion rod has been knocked.
Wait until the message disappears, then lower the rod.

Update Vitrex

Error 13

BASIC has a newer SW then VITREX
Call service

Force Sensor

Error 15

Forcesensor defective.
Call service

Adjust Venturi

Error 16

Impossible venturi sensor calibration values.
Call service.

Adjust Force Sens

Error 17

Impossible peristaltic sensor calibration values.
Call service.

Check Venturi

Error 18

Venturi System: Vacuum deviation to high.
Call service

Errors 5, 6 and 7 do not appear for this unit

Summary of Contents for OS3

Page 1: ...29 05 09 VV016011E 1 Operating Instructions OS3 Base unit VC830100 VV016011E Phaco Emulsification System 0297...

Page 2: ...taltic operation 13 8 2 3 Infusion pressure IOP 13 8 2 4 Continuous irrigation 14 8 2 5 Reflux 14 8 2 6 Proportional Override 14 8 3 Ultrasonic phaco 14 8 3 1 Phaco test 14 8 3 2 Phaco operation 14 8...

Page 3: ...may be impaired or the pacemaker itself damaged If in doubt the cardiology department must be consulted 2 The functioning of other electro medical equipment may be affected during the operation of bip...

Page 4: ...k green arrow buttons immediately beneath the display The lower half of the control panel houses the displays and adjustment buttons for the irrigation aspiration system The lowest row on the control...

Page 5: ...g factor can be changed see 8 3 4 PULSE button Switches on pulse or burst operation for phaco 1 2 3 and switches it off when the button is pressed again If the button is held down the pulse or burst p...

Page 6: ...tween positions 0 1 2 and 3 If the pedal is pressed slowly the settings 0 3 can be felt as a slight resistance This resistance can be changed in ParaProg low medium high The pedal is brought to positi...

Page 7: ...ear Linear instrument control Aspiraton Phaco VIT DIA LOW linear CAPS REGULAR fix VISCO INJECTION linear Reflux Instant diathermy Pulse on off linear phaco Pos 3 Instrument Phaco VIT DIA CAPS Override...

Page 8: ...the installed software appears in weakly illuminated lettering The unit is now ready and can be prepared for the operation 5 1 Replacing fuses Press the clip on fuse holder P to the right until it pop...

Page 9: ...trument connection tubes must be kept in the sterile area The sterile person must not insert the tubes into the unit Turn the cassette rotary switch in an anti clockwise direction and open the cassett...

Page 10: ...P button twice 6 3 Removing and cleaning the cassette 6 3 1 Cassettes for single use Fig 3 4 Remove the tubes from the instruments and join together connections 1 and 2 to prevent the liquid from drip...

Page 11: ...g the irrigation tube i and the aspiration tube a into the I A handpiece The luer connectors are designed so that it is impossible to connect the instruments wrongly Attach the chosen tip Wet the tip...

Page 12: ...CO1 PHACO2 PHACO3 CMP1 CMP2 CMP3 as well as VIT A VIT B and VIT PN or VIT A1 VIT A2 etc Calling up values Press the above buttons briefly to call up the stored values immediately Storing new values Pr...

Page 13: ...03 or VV 630010 are being used you can change between the venturi and peristaltic pump at any time To do so the pedal must be in the zero position Effect the change by pressing the venturi button on t...

Page 14: ...be tested before beginning the operation To do so fit a filled test chamber to the handpiece and press the TEST button Never carry out the TEST procedure whilst the handpiece is in the eye NOTE The te...

Page 15: ...ional to the pedal deflection Note CMP1 2 3 have their own memories By holding down the corresponding button phaco 1 2 3 see 8 1 for a short period all settings will be memorized 8 3 5 Occlusion Mode...

Page 16: ...ains available without having to set the unit to diathermy mode by pressing the DIA button This function can be set to DiaLOW or DiaHIGH or switched off in ParaProg NOTE ParaProg can be used to determ...

Page 17: ...nd 3 Duallinear vitrectomy This operating mode enables separate control of the pump by vertical pedal movement and control of the cutting rate by moving the pedal horizontally to the left or vice vers...

Page 18: ...es Every surgeon develops his own preferred operating technique which also requires corresponding specific setting values for the various stages of the operation The OS3 unit can optimally accommodate...

Page 19: ...are temperature 134 C 138 C min cycle time 3 minutes When the instruments are removed from the sterilisation system they must be cooled to room temperature before commencing an operation The instrume...

Page 20: ...itrectomy 200Vss at instrument input Aspiration flow 0 50 ml min 25 Vacuum 0 600 mmHG 10 0 80kPa 10 Noise level 70 dB A Dimensions 380 x 150 x 340 mm W H D Weight Unit 14 8 kg Pedal 4 8 kg Transport a...

Page 21: ...heel System ready The self test has been completed successfully Check tip The tip is not properly attached See Section 8 3 1 Call Service Error 1 Contact service Internal voltages defective The unit c...

Page 22: ...aco accessories included Vitrectomy accessories included 18 Calibration and maintenance Provided that the cassettes are replaced regularly in accordance with the instructions in the packaging this uni...

Page 23: ...nnection NIST EN 397 L Slot for smart card see service manual M1 Pedal port M2 Reserve port M3 Control panel port N Serial interface see service manual O Mains connection socket P Fuse holder Q Mains...

Reviews: