Archive OS3 Operating Instructions Manual Download Page 17

29.05.09

VV016011E

17

8.6

Vitrectomy

Press one of the buttons VIT A, VIT B or VIT PN.

VIT A1

1200/min

Surgeon 2

IOP 75cm

10ml/min 250mmHg

The display fields indicate the function selected, together with
the values last used for cutting rate, flow (Venturi rate) and
suction. The arrow buttons can be used to increase or reduce
these values at any time.

Unless otherwise specified in ParaProg, the VIT A connection
serves to operate an SDS or SUS guillotine vitrectomy
stripper, and the VIT B connection the micro-scissors.

The PN connection is for operating the pneumatically-driven
Twinac vitrectomy instrument.

NOTE: In ParaProg, VIT A, VIT B and VIT PN can be set up for stripper or
scissor operation, according to preference.

Vitrectomy multimodulation programs

Three multimode memories are available for each of the
functions VIT A, VIT B, VIT PN. They are identified as VIT
A1, VIT A2, etc. on the function display. Switching between
these multimode programs is accomplished by repeated
depressing of the corresponding function key (VIT A, VIT B,
VIT PN) or by swinging the pedal to the right while in position
0 or 1.

NOTE: in ParaProg you can barr these multimode memories or limit to two
memories only.

Storing of the values for the vitrectomy multimode programs
is done as described in Section 8.1.

TEST function

Press the TEST button to check whether the connected
instrument is functioning properly (operation with the set
number of cutting movements).

The surgeon controls the vitrectomy procedure by using the
pedal. When the pedal is active, the value display fields show
the current values. Unless otherwise specified in ParaProg,
the following controls are available ("linear vitrectomy"):

Vertical step 1:

Irrigation

Vertical step 2:

Linear aspiration

Vertical step 3:

Linear cutting

Horizontal left:

Instant diathermy (only in position 0)

Horizontal right: Single cut (stripper VIT A) Close scissors

(VIT B)

NOTE:

In ParaProg, you can specify the reverse sequence: position 2 cut,

position 3 aspiration, and program the horizontal right movement to switch
between VIT A1, VIT A2 etc. Instant diathermy can also be made available for
operation in positions 2 and 3.

Duallinear vitrectomy

This operating mode enables separate control of the pump by
vertical pedal movement and control of the cutting rate by
moving the pedal horizontally to the left (or vice versa). It is
programmed in ParaProg.

To exit the VIT function, you can either press a new function
button or the RESET button.

As long as the pedal is active (positions 2, 3), only the arrow
buttons and the RESET, AUX and VOICE buttons will
respond.

9

System communication

9.1

Visual displays

Light indicators serve to show selected, programmed and
current values and important information about all other unit
states on the control panel and remote controller. Warnings
and instructions are displayed in the language selected in
ParaProg. See also
Section 17.

9.2

Acoustic signals

The unit uses acoustic signals to indicate the state of the I/A
system and the output values. Different tones represent the
various states:

INFUSION OPEN
Sound of a slow drip
ASPIRATION ACTIVE
Sound of a rapid drip, whose pitch rises as the vacuum level
increases.
VACUUM LIMIT REACHED
Short beeps
REFLUX (pump reversed)
A high-pitched, rapidly-repeating signal
DIATHERMY ACTIVE
High-pitched sequence of sounds, rising with increased
output
HF CUT ACTIVE
Low-pitched sequence of tones (IDK)
Medium-pitched sequence of tones (STT)
High-pitched sequence of tones (CAPSULOTOMY)
PHACO ACTIVE 1)
Repeating combination of three beeps

Volume control
Hold down the AUX button, and select a volume level
between 0-100% with the "SOUND" arrow buttons.

NOTE:

The CONTIN option can be selected in ParaProg for the acoustic

monitoring of the vacuum. The vacuum is then indicated by a wailing tone,
which rises as the vacuum value climbs to the set maximum value.
1)

If specified in ParaProg

9.3

Voice confirmation

Voice confirmation will notify you about the selected function
and values as well as about warnings and instructions. You
can set up the scope of voice information for each surgeon in
ParaProg, and switch each of the following messages
individually to silent:

Warnings and instructions (VOICE MESSG)
Confirmation of selected functions (VOICE MODES)
Confirmation of set values (VOICE VALUES)
Confirmation only of changed values (CHANGE)

In addition, you can specify whether the messages should be
given out automatically (VOICE AUTOM) or only when the
VOICE button is pressed.

Volume control

Summary of Contents for OS3

Page 1: ...29 05 09 VV016011E 1 Operating Instructions OS3 Base unit VC830100 VV016011E Phaco Emulsification System 0297...

Page 2: ...taltic operation 13 8 2 3 Infusion pressure IOP 13 8 2 4 Continuous irrigation 14 8 2 5 Reflux 14 8 2 6 Proportional Override 14 8 3 Ultrasonic phaco 14 8 3 1 Phaco test 14 8 3 2 Phaco operation 14 8...

Page 3: ...may be impaired or the pacemaker itself damaged If in doubt the cardiology department must be consulted 2 The functioning of other electro medical equipment may be affected during the operation of bip...

Page 4: ...k green arrow buttons immediately beneath the display The lower half of the control panel houses the displays and adjustment buttons for the irrigation aspiration system The lowest row on the control...

Page 5: ...g factor can be changed see 8 3 4 PULSE button Switches on pulse or burst operation for phaco 1 2 3 and switches it off when the button is pressed again If the button is held down the pulse or burst p...

Page 6: ...tween positions 0 1 2 and 3 If the pedal is pressed slowly the settings 0 3 can be felt as a slight resistance This resistance can be changed in ParaProg low medium high The pedal is brought to positi...

Page 7: ...ear Linear instrument control Aspiraton Phaco VIT DIA LOW linear CAPS REGULAR fix VISCO INJECTION linear Reflux Instant diathermy Pulse on off linear phaco Pos 3 Instrument Phaco VIT DIA CAPS Override...

Page 8: ...the installed software appears in weakly illuminated lettering The unit is now ready and can be prepared for the operation 5 1 Replacing fuses Press the clip on fuse holder P to the right until it pop...

Page 9: ...trument connection tubes must be kept in the sterile area The sterile person must not insert the tubes into the unit Turn the cassette rotary switch in an anti clockwise direction and open the cassett...

Page 10: ...P button twice 6 3 Removing and cleaning the cassette 6 3 1 Cassettes for single use Fig 3 4 Remove the tubes from the instruments and join together connections 1 and 2 to prevent the liquid from drip...

Page 11: ...g the irrigation tube i and the aspiration tube a into the I A handpiece The luer connectors are designed so that it is impossible to connect the instruments wrongly Attach the chosen tip Wet the tip...

Page 12: ...CO1 PHACO2 PHACO3 CMP1 CMP2 CMP3 as well as VIT A VIT B and VIT PN or VIT A1 VIT A2 etc Calling up values Press the above buttons briefly to call up the stored values immediately Storing new values Pr...

Page 13: ...03 or VV 630010 are being used you can change between the venturi and peristaltic pump at any time To do so the pedal must be in the zero position Effect the change by pressing the venturi button on t...

Page 14: ...be tested before beginning the operation To do so fit a filled test chamber to the handpiece and press the TEST button Never carry out the TEST procedure whilst the handpiece is in the eye NOTE The te...

Page 15: ...ional to the pedal deflection Note CMP1 2 3 have their own memories By holding down the corresponding button phaco 1 2 3 see 8 1 for a short period all settings will be memorized 8 3 5 Occlusion Mode...

Page 16: ...ains available without having to set the unit to diathermy mode by pressing the DIA button This function can be set to DiaLOW or DiaHIGH or switched off in ParaProg NOTE ParaProg can be used to determ...

Page 17: ...nd 3 Duallinear vitrectomy This operating mode enables separate control of the pump by vertical pedal movement and control of the cutting rate by moving the pedal horizontally to the left or vice vers...

Page 18: ...es Every surgeon develops his own preferred operating technique which also requires corresponding specific setting values for the various stages of the operation The OS3 unit can optimally accommodate...

Page 19: ...are temperature 134 C 138 C min cycle time 3 minutes When the instruments are removed from the sterilisation system they must be cooled to room temperature before commencing an operation The instrume...

Page 20: ...itrectomy 200Vss at instrument input Aspiration flow 0 50 ml min 25 Vacuum 0 600 mmHG 10 0 80kPa 10 Noise level 70 dB A Dimensions 380 x 150 x 340 mm W H D Weight Unit 14 8 kg Pedal 4 8 kg Transport a...

Page 21: ...heel System ready The self test has been completed successfully Check tip The tip is not properly attached See Section 8 3 1 Call Service Error 1 Contact service Internal voltages defective The unit c...

Page 22: ...aco accessories included Vitrectomy accessories included 18 Calibration and maintenance Provided that the cassettes are replaced regularly in accordance with the instructions in the packaging this uni...

Page 23: ...nnection NIST EN 397 L Slot for smart card see service manual M1 Pedal port M2 Reserve port M3 Control panel port N Serial interface see service manual O Mains connection socket P Fuse holder Q Mains...

Reviews: