Cleaning and sterilisation regulations

IMPORTANT!

Make sure to adhere to guideline TN999042 (as well as to all

of the other relevant requirements that have been made

available) when preparing the instruments for re-use!

12.1 Cleaning the Control Device

The device and the control panels should be cleaned at the
end of each day of operation. Remove all BSS residue using
a soft fiber-free cloth that has been dampened with medicinal
benzene. Make sure to keep the control panels and the
device dry at all times. They are not to be spayed or rinsed.

12.2 Cleaning the Instruments

Immediately after the operation, immerse the diathermy tips,
cutting heads and handpieces in BSS or distilled water and
rinse thoroughly. The cleaning instructions supplied with the
instruments should be strictly observed!

Only use distilled or de-ionised water, neutral cleaning agents
and a soft lint-free cloth or soft sponge. Ensure that all
instruments are free from blood, tissue and impurities caused
by saline deposits or other substances. Rinse carefully and
thoroughly with distilled water, and clean carefully with
compressed air. Do not use oxygen or any other gases!

12.3 Sterilisation

IMPORTANT!

Clean diathermy and capsulotomy tips immediately after

removal from the eye and keep moist with BSS or distilled

water until they can undergo final cleaning.

Steam sterilisation is the recommended method for tips,
handpieces, instruments and re-usable cassettes. ETO
sterilisation is not recommended, and gamma sterilisation is
unacceptable because of the instability of the materials.

IMPORTANT!

The user is responsible for using the correct sterilisation

method, including taking any precautions to ensure

bacteriological safety.

After cleaning (as described in Section 12.1), tips,
handpieces and instruments must be sterilised in the
autoclave with a supporting air removal device. The
recommended values are: temperature 134°C…138°C, min.
cycle time 3 minutes.

When the instruments are removed from the sterilisation
system, they must be cooled to room temperature before
commencing an operation.

The instruments should be dismantled into their individual
parts before being sterilised.

IMPORTANT!

The instruments must be sterilised before every use!

13

Accessories and replacement parts

Unit accessories

VC830200

Vitrex add-on module for vitreo-retinal surgery

VE830001

Unit trolley with infusion bar drive

VE830010

Programmable duallinear pedal

VE830020

Remote controller with illumination

VE830025

Second control panel for base unit

VX520010

3.15 AT fuses, high breaking capacity

Consumable materials

VV630003

Twin Vac cassette, autoclavable (pack of 3)

VV630010

Twin Vac cassette, single use (pack of 10)

VV635010

Venturi cassette with second aspiration port,
single use, (pack of 10)

VV630011A

Phaco Pack (pack of 10)

Phaco instruments

VG800011

Ultrasonic phaco handpiece

VV800415

Phaco tip, "Excellerator 2", titanium, 15°

VV800430

Phaco tip, "Excellerator 2", titanium, 30°,

VV800530

Phaco tip, "Ergo Excellerator", titanium, 30°
(Kelman)

VV800050

Pars Plana Phaco tip, titanium

Vitrectomy instruments

VE100100

Drive for SDS instruments

VV101301

SDS guillotine cutting head

VV101201

SDS Klöti cutting head

VV101501

SDS micro scissors

VG601151

SDS irrigation sleeve, 1.5 mm

VE103100

Drive for SUS instruments

VV103006

SUS guillotine cutting head, single use,
6 pieces

VV104010

Twinac pneumatic cutter, single use, box of 10

VG601351

SUS and Twinac irrigation sleeve, 1.5mm,
reusable.

Diathermy instruments

VE201712

"Plug-on" diathermy handpiece

VE201722

"Plug-on" eraser tip

VE201723

"Plug-on" endo diathermy tip, 0.89 mm

VE201724

"Plug-on" endo diathermy tip, 0.45 mm

VE201726

"Plug-on" capsulotomy tip

VE203902

"Plug-on" bipolar forceps

14

Authorised service centres

Switzerland

Oertli Instrumente AG

(manufacturer)

Hafnerwisenstr. 4
CH-9442 Berneck
Tel. +41-71-7474200

Information about other service centres can be obtained from
the manufacturer.

Authorised representative in the EU
Germany

Oertli Instrumente GmbH

Magnolienweg 14
D-63741 Aschaffenburg

Summary of Contents for OS3

Page 1: ...29 05 09 VV016011E 1 Operating Instructions OS3 Base unit VC830100 VV016011E Phaco Emulsification System 0297...

Page 2: ...taltic operation 13 8 2 3 Infusion pressure IOP 13 8 2 4 Continuous irrigation 14 8 2 5 Reflux 14 8 2 6 Proportional Override 14 8 3 Ultrasonic phaco 14 8 3 1 Phaco test 14 8 3 2 Phaco operation 14 8...

Page 3: ...may be impaired or the pacemaker itself damaged If in doubt the cardiology department must be consulted 2 The functioning of other electro medical equipment may be affected during the operation of bip...

Page 4: ...k green arrow buttons immediately beneath the display The lower half of the control panel houses the displays and adjustment buttons for the irrigation aspiration system The lowest row on the control...

Page 5: ...g factor can be changed see 8 3 4 PULSE button Switches on pulse or burst operation for phaco 1 2 3 and switches it off when the button is pressed again If the button is held down the pulse or burst p...

Page 6: ...tween positions 0 1 2 and 3 If the pedal is pressed slowly the settings 0 3 can be felt as a slight resistance This resistance can be changed in ParaProg low medium high The pedal is brought to positi...

Page 7: ...ear Linear instrument control Aspiraton Phaco VIT DIA LOW linear CAPS REGULAR fix VISCO INJECTION linear Reflux Instant diathermy Pulse on off linear phaco Pos 3 Instrument Phaco VIT DIA CAPS Override...

Page 8: ...the installed software appears in weakly illuminated lettering The unit is now ready and can be prepared for the operation 5 1 Replacing fuses Press the clip on fuse holder P to the right until it pop...

Page 9: ...trument connection tubes must be kept in the sterile area The sterile person must not insert the tubes into the unit Turn the cassette rotary switch in an anti clockwise direction and open the cassett...

Page 10: ...P button twice 6 3 Removing and cleaning the cassette 6 3 1 Cassettes for single use Fig 3 4 Remove the tubes from the instruments and join together connections 1 and 2 to prevent the liquid from drip...

Page 11: ...g the irrigation tube i and the aspiration tube a into the I A handpiece The luer connectors are designed so that it is impossible to connect the instruments wrongly Attach the chosen tip Wet the tip...

Page 12: ...CO1 PHACO2 PHACO3 CMP1 CMP2 CMP3 as well as VIT A VIT B and VIT PN or VIT A1 VIT A2 etc Calling up values Press the above buttons briefly to call up the stored values immediately Storing new values Pr...

Page 13: ...03 or VV 630010 are being used you can change between the venturi and peristaltic pump at any time To do so the pedal must be in the zero position Effect the change by pressing the venturi button on t...

Page 14: ...be tested before beginning the operation To do so fit a filled test chamber to the handpiece and press the TEST button Never carry out the TEST procedure whilst the handpiece is in the eye NOTE The te...

Page 15: ...ional to the pedal deflection Note CMP1 2 3 have their own memories By holding down the corresponding button phaco 1 2 3 see 8 1 for a short period all settings will be memorized 8 3 5 Occlusion Mode...

Page 16: ...ains available without having to set the unit to diathermy mode by pressing the DIA button This function can be set to DiaLOW or DiaHIGH or switched off in ParaProg NOTE ParaProg can be used to determ...

Page 17: ...nd 3 Duallinear vitrectomy This operating mode enables separate control of the pump by vertical pedal movement and control of the cutting rate by moving the pedal horizontally to the left or vice vers...

Page 18: ...es Every surgeon develops his own preferred operating technique which also requires corresponding specific setting values for the various stages of the operation The OS3 unit can optimally accommodate...

Page 19: ...are temperature 134 C 138 C min cycle time 3 minutes When the instruments are removed from the sterilisation system they must be cooled to room temperature before commencing an operation The instrume...

Page 20: ...itrectomy 200Vss at instrument input Aspiration flow 0 50 ml min 25 Vacuum 0 600 mmHG 10 0 80kPa 10 Noise level 70 dB A Dimensions 380 x 150 x 340 mm W H D Weight Unit 14 8 kg Pedal 4 8 kg Transport a...

Page 21: ...heel System ready The self test has been completed successfully Check tip The tip is not properly attached See Section 8 3 1 Call Service Error 1 Contact service Internal voltages defective The unit c...

Page 22: ...aco accessories included Vitrectomy accessories included 18 Calibration and maintenance Provided that the cassettes are replaced regularly in accordance with the instructions in the packaging this uni...

Page 23: ...nnection NIST EN 397 L Slot for smart card see service manual M1 Pedal port M2 Reserve port M3 Control panel port N Serial interface see service manual O Mains connection socket P Fuse holder Q Mains...

Reviews: