29.05.09
VV016011E
20
15
Technical data
Supply pressure*
Air 6.5…10 bar, max. 40 l/min,
NIST EN-739 connection
Mains voltage
100…240 VAC
Mains frequency
50…60 Hz
Power consumption
270 VA (590 VA with Vitrex)
Fuses
3.15 AT, high breaking capacity
Operating mode
continuous
Application parts
non-earthed, type BF (IEC 601),
Exception: Phaco type B
Protection class
I
CE classification
2B
HF output power
Phaco: 26....30kHz 28kHz nom.
46W ± 8 W (100%/1000Ω)
Diathermy
500 kHz 0-8 watt, nom. at 50 Ohm
Capsulotomy
automatic energy
500kHz control, regular 6.5W
high 9.9W, eff. at 50Ω
IDK 5.5W
STT 6W at 50 Ω
Vitrectomy
VIT A, B: 30 – 1200/min
VIP PN: 1500/min
optionally 3000/min
Max. HF output voltage Diathermy: 60Vss (47Ω/100%)
HF CUT: 270Vss (47Ω)
Phaco: 550Vss (1100Ω/100%)
Vitrectomy: 200Vss (at instrument
input)
Aspiration flow
0-50 ml/min ±25%
Vacuum
0-600 mmHG ±10%
0-80kPa ±10%
Noise level
<70 dB (A)
Dimensions
380 x 150 x 340 mm (W/H/D)
Weight
Unit 14.8 kg
Pedal 4.8 kg
Transport and storage
Temperature: -20°C…+55°C
Conditions
Air pressure: 500hPh…1060hPa
Relative air humidity: 10%…95%,
non condensing
Operating conditions
Temperature:10°C...40°C
Air pressure: 700...1060hPa
Relative air humidity: 20%....80%
non condensing
* If there is no pressure connection, the venturi pump and VIT PN cannot be
operated.
HF-CUT
0
2
4
6
8
10
12
0
100
200
300
400
500
600
700
800
900
1000
Load resistance [Ohm]
O
u
tp
u
t
p
o
w
e
r
[W
]
CAPSULE REGULAR
CAPSULE HIGH
GLAUKOMA IDK
GLAUKOMA STT
Diathermy
0
1
2
3
4
5
6
7
8
9
0
100
200
300
400
500
600
700
800
900
1000
Load resistance [Ohm]
O
u
tp
u
t
p
o
w
e
r
[W
]
DIA 5%
DIA 25%
DIA 50%
DIA 75%
DIA 100%
DIA 50 Ohm and 100 Ohm
0
1
2
3
4
5
6
7
8
9
0
10
20
30
40
50
60
70
80
90
100
Set value [%]
O
u
tp
u
t
[W
]
DIA 50 Ohm
DIA 100 Ohm