X-Ray & ABS
20
2. Put a probe inside the generator on the shunt SH3 (left side of the resistor or the big yellow capacitor in parallel, C8 of kV
Measure PCB on HV Tank (2).
3. Connect a second probe on the mechanical ground (3).
4. Check the resistor value with a multi-meter. It should be 5 ohms (± 0.05).
X-Ray Stator Test
1. Apply power to the system.
Caution:
Do not operate the system if the stator motor cannot be heard and is not operating.
2. Verify that the stator motor is rotating by listening for the “whirring” sound coming from the X-ray tube head.
3. Verify that the control Panel does not display the error message “Stator not on.” If the message is displayed, refer to the
Diagnostics section of this manual and correct the problem.
kVp/mA Servo Loop Test
WARNING: Steps within this procedure produce x-rays. Use appropriate precautions.
1. Select MANUAL FLUORO mode and set the technique for 40 kVp @ 4 mA.
2. Verify that the kVp and mA can be controlled using the control panel switches.
3. Press the X-RAY ON switch and verify that no error messages are displayed on the Control Panel Display. If an
error message is displayed, go to the Diagnostics section of this manual and correct the fault.