Disassembly
The Invasive mA Measure can be made only on the left side of the Inverter Unit. (Refer to the
Generator Assembly Diagram
in
this section.)
1. Unscrew the screws (see note below) securing the two EMC cover panels (1) on the HV connections side and remove them.
Note:
Do not to let screws fall into the HV connector receptacles. Cover the receptacles
with plastic caps.