![Archive OEC UroView 2800 Manual Download Page 31](http://html.mh-extra.com/html/archive/oec-uroview-2800/oec-uroview-2800_manual_2964230031.webp)
X-Ray & ABS
31
D
S1
00
D
S1
02
D
S2
00
D
S1
01
DS100 -Yellow-Low IGBT*
(Q100) Gate Command running
DS101 - Green - Presence of
+20 V Supply on low IGBT*
Gate Command
DS200 - Yellow - High
IGBT* (Q200) Gate
Command running
DS102 - Green - Presence
of -10 V Supply on low
IGBT* Gate Command
DS202 - Green - Presence
of -10 V Supply on high
IGBT* Gate Command
DS300 - Neon (Orange) -
Presence of voltage on DC
bus for Flyback Converter to
create power supplies for
both Gate Commands
* Insulated Gate
Bipolar Transistor
D
S3
00
D
S2
01
D
S2
02
DS201 - Green - Presence
of +20 V Supply on high
IGBT* Gate Command
DC bus
Inverter
current
feedback
ILP
ILR
Parallel inductor
temperature
DS200 flashes during
pulsed fluoro; is on steady
for regular X-ray (Fluoro).
DS100 – On
during X-ray
(Fluoro)
Summary of Contents for OEC UroView 2800
Page 19: ...X Ray ABS 19 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation ...
Page 37: ...X Ray ABS 37 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation ...