X-Ray & ABS
12
ABS Control during Pulsed Boost Fluoro Mode
Function Description
kVp
kVp frozen at its current value by software.
mA
mA is incremented until the video level equals the video level index.
Camera Gain
Camera Gain is reduced, and the ABS system adjusts mA to achieve
correct video level. If the correct video level is not reached before the
mA limit is reached, the camera gain is increased.
Camera Iris
Iris fully closed.
ABS Control during Boost Fluoro Mode
Function Description
kVp
kVp frozen at its current value by software.
mA
mA is incremented until the video level equals the video level index.
Camera Gain
Camera Gain is reduced, and the ABS system adjusts mA to achieve
correct video level. If the correct video level is not reached before the
mA limit is reached, the camera gain is increased.
Camera Iris
Iris fully open.
Pulse
Pulsed X-rays can be used to reduce the total patient radiation exposure, but are not used with film or digital spot mode. A preset
number of X-rays are generated while the Pulse button is pressed. Pulsed X-rays are available to any Fluoro mode, whether
Auto or Manual Fluoro.