![AR 10046836 Operating And Service Manual Download Page 46](http://html1.mh-extra.com/html/ar/10046836/10046836_operating-and-service-manual_2962294046.webp)
Model 500W1000C
26
Rev A
2.6.6.16
AC Power-On Defaults
Default settings that are applied at AC mains power-on can be changed by adding the following prefix to select
commands.
Syntax:
DEFAULT:
Compatible commands:
Level Adjust
LEVEL:GAIN
LEVEL:DET
(Not available on all models)
LEVEL:THR
(Not available on all models)
LEVEL:RESP
(Not available on all models)
Mode Select
MODE:MANUAL
(Not available on all models)
MODE:PULSE
(Not available on all models)
MODE:ALC<space>INT
(Not available on all models)
MODE:ALC<space>EXT
(Not available on all models)
NOTES:
1.
Use the command
DEFAULT:FACTORY
to reset all applicable settings back to their factory defaults.
2.
All applicable defaults can be queried, except
DEFAULT:FACTORY
, by adding a ? character in place of
the setting parameter.
3.
If the ALC Lockout Feature (not available on all models) is engaged, the default ALC values set with this
command will not be used.
Example 1:
To set the default RF Gain to 75%, send the following command:
DEFAULT:LEVEL:GAIN75<LF>
Example 2:
To query the default RF Gain setting, send the following command:
DEFAULT:LEVEL:GAIN?<LF>
Response:
DEFAULT:LEVEL:GAIN75<LF>
Example 3:
To set the default mode to manual, send the following command:
DEFAULT:MODE:MANUAL<LF>