![AR 10046836 Operating And Service Manual Download Page 43](http://html1.mh-extra.com/html/ar/10046836/10046836_operating-and-service-manual_2962294043.webp)
Model 500W1000C
Rev A
23
Example:
To read the state, send the following query.
STATE?<LF>
Response:
STATE=<space>8301<LF>
(Remote Mode, Power On, RF OFF, and Manual Mode)
2.6.6.10
Forward Power
Query to get the forward power.
Syntax:
FPOW?
Parameters:
None
Response Format:
FPOW=x<LF>
Where:
x
= 0 to 99999
Values are corrected and linearized. They can be up to five digits in length. Leading zeros
are read as spaces. Units are Watts.
Example:
To find out the forward power, send the following query.
FPOW?<LF>
Response:
FPOW=<space><space><space>54<LF>
(54 Watts of forward power)
2.6.6.11
Reverse Power
Query to get the reverse power.
Syntax:
RPOW?
Parameters:
None
Response Format:
RPOW=x<LF>
Where:
x
= 0 to 99999
Values are corrected and linearized. They can be up to five digits in length. Leading zeros
are read as spaces. Units are Watts.
Example:
To find out the reverse power, send the following query.
RPOW?<LF>
Response:
RPOW=<space><space><space><space>9<LF>
(9 Watts of reverse power)