background image

 
 

Maintenance/Checklist: 

To achieve the best counting result, be content with the maintenance of the counter. The units should 
always be cleaned after use. Follow also this checklist before every use of the counter: 
Use a moist cloth to clean the transparent pipe in the counter. Open the inspection-hatch and check 
that everything is clean: window in front of lenses, transparent pipe and the light source. (The lenses 
look through the top of the pipe towards the light source at the bottom of the pipe). Remove any 
droplets on the window and clean it with DPI-MICROCLEAN. (We want to achieve the antidroplet 
effect – opposite effect to car polish/wax). 
 
During counting, drops of water may form at the window in front of the lenses or on the splash-cover. 
Drops of water on the window/splash-cover can lead to inaccurate count typically if the fish are small. 
To avoid this problem, clean the window with DPI-MICROCLEAN (the counter must be stopped). 
Keep the splash-cover open if the fish is smaller than 200g. During the counting operation, check that 
there is no droplets on the window. 
 

AquaScan counting principle and capacity: 

It is of no matter whether one or several fish pass the sensor 
simultaneously. AquaScan measures the size (shape) of all 
objects passing through the unit. Capacity is only limited that the 
fish must not lie on top of each other and then cover the view of 
the sensor (lens). The number calculated is the total size of all 
fishes divided on the last calculated average size. The average 
size is continuously calculated and updated during the whole 
counting process. (AquaScan will calculate the average size 
when a “single fish” is registrated). You may recognize that 
sometimes AquaScan does not count when a fish pass through or 
count two instead of one etc. In the first case the fish may be too small to initiate a count or the 
average has been changed. However, the size is memorized and this will have an effect on the total 
count. 
 

Calibration: 

When the operator types in the average weight, at least the first 125 fish passing through this 
Registration Unit will be used to calibrate the system for this channel. The text 

Cal

 will follow the 

number of group of fish (one or several fish) in channels where calibration is in progress. During this 
mode the system will not separate between one or several fish passing through at the same time. 
 
During calibration, at least 50 of the first 125 group of fish in the channel need to be a single fish 
(distance between every fish). AquaScan will show the correct number of fish passed during the 
calibration as soon as the calibration is finished. When at least 125 fish have been recorded, the 
system will automatically change calibration mode to normal counting. 
 
Ideally, if all measurements are OK you should see 

124 Cal

125 Cal

 (one channel viewed) and then 

the correct number e.g. 

145

 without the text 

Cal

. If some of the measurements during calibration are 

bad (you will see 

126 Cal

127 Cal

, ....) some of the fish will not be used to calibrate the system. Bad 

measurements may be caused by e.g. too slow speed of the fish or too much water. 
 
 

Page 15 of 20 

Summary of Contents for CSE Series

Page 1: ...AquaScan Manual Rev 1 6 KE CSE series AquaScan User s Guide 4 03 Model KE series AquaScan General Directions version 1 6 Model CSE series Copyright AquaScan 2010 Page 1 of 20...

Page 2: ...sults menu 10 Results menu menu page 1 10 1 Res PC 10 2 Res DISP 10 Results menu menu page 2 12 AquaScan General directions 13 Electrical connection disconnection 13 Connection of transportation pipe...

Page 3: ...uaScan system AquaScan General Directions Contains vital information on the arrangement connection and counting principle of the AquaScan system It is very important that operators read General Direct...

Page 4: ...ing will be showed when the program transfer is finished x may be 1 2 3 or 4 C1 C2 C3 and C4 are related to the inputs at the Control Unit For each channel type the average weight in gram for the fish...

Page 5: ...g This means that calibration is at least in progress for channel 1 If the same indicator C3 is pressed again the display will return to show the sum of units C1 and C3 Now a press on indicator C1 hi...

Page 6: ...eed for Cx Ne in for Cx Nst in for Cx Number for Cx If two Registration Units are connected data for both can be displayed at the same time Number for Cx Number for Cx in related to total Number for C...

Page 7: ...sume counting x may be 1 2 3 or 4 This is a warning counting is not stopped The speed calculations in the Registration Unit are having a problem This may be caused by too much water in a dry counter t...

Page 8: ...e time of day submenu Clock menu time of day submenu Yes selects time of day adjustment No or Stop goes back to top level menu If Yes is selected the hour is first set by two digits and Enter then min...

Page 9: ...t to 950 the displayed count for C1 will be 95 of the value compared to if the weight factor was 1000 In the same manner if the Count factor for C2 is set to the displayed count for C2 will be 5 great...

Page 10: ...hannel if more channels was in use v or displays next or previous screen Stop goes back to Results menu Number of fishes for selected channel tells the percentage of the total number of fishes for all...

Page 11: ...be greater than 2 m s Max 3 m s for fish less than 200g Recommended speed is at least 3m s for bigger fish Display counting result Ne Nst screen Calib Weight 25g C1 05 10 06 09 49 01 Avg Weight 24g C1...

Page 12: ...ting was stopped Display counting result Run time screen How long the counting did take Display counting result Effectiveness screen Number of kg hour for displayed channel Results menu menu page 2 Ru...

Page 13: ...female connectors when not in use to avoid access of dirt and water Connection of transportation pipe to Reg Unit The Registration Unit should be installed in such a way that the long sides are vertic...

Page 14: ...ation Unit and pipes should be tilted forward to a suitable angle min 25 degrees to assure steady flow approx 2 5 3 meter second through the counter The Control Unit can view the average speed of the...

Page 15: ...n the last calculated average size The average size is continuously calculated and updated during the whole counting process AquaScan will calculate the average size when a single fish is registrated...

Page 16: ...m the number of fish calculated during the previous counting operation Details about the rub window alarm If the screen reads RUB WINDOW C2 no counting will be performed on channel 2 and you will loos...

Page 17: ...e Control Unit responds to other equipment The arrow buttons may be used to navigate between parameters in each code set To move to the next parameter in a code set press 6 key or N To move to the pre...

Page 18: ...PC connection LOG 9 REF 50 ALD 10 CSE1600 ALD 5 VEL 30 DLL 50 DLH 50 IRM 0 CON 0 WGT Constantly changing WLL 30 WLH 30 Parameter setting 2 Delete 4 NMS 125 NOL 70 MAT 512 AMI 1920 DSP 100 STP 2 SUN 40...

Page 19: ...Page 19 of 20...

Page 20: ...Page 20 of 20...

Reviews: