15
Set Time and Date Screen
The current time, day, week, and date are important for managing
the treatment program and timing of ORP, biocide control, and for
time stamping the data log. Only two values are adjustable but each
contains several components.
CLK denotes clock time in 24–hour format. Hours and minutes are
set separately.
DATE denotes the current date. Day, month, and year are set
separately. This screen also displays the day of the week which
it calculates from the date and the week number according to the
following outline. The week is determined by the controller and
changes at midnight between Saturday and Sunday. Week is used
in the set up of biocide dosing programs.
Set Data Logger Screen
This screen is used to set the data logging options. The options are
shown below.
Log interval: [30 mins]
Options: 1, 2, 5, 10, 15, 30, 60, 120, 240 (Expressed in minutes).
Use the arrow keys to cycle these selections.
Log period - is a calculated value that the controller computes and
is expressed in days. The data log file will hold approximately 2400
entries which means that the duration is a function of the log interval.
E.g. a log interval of 30 minutes will produce a total log period of 35
days.
Set Alarms Screen
Allows the user to configure various alarm functions within the
controller.
IGNORE FLOW:
[OFF]
When set to ON disables loss of flow alarm.
Options:
OFF/ON
Use arrow keys to cycle these selections.
REPEAT:
[Every Morning]
Options: Every Morning/Every 4 hrs/Once only/Never.
Use arrow keys to cycle these selections.
IGNORE CAL:
[ON]
When set to OFF enables out of calibration’ alarms.
Options:
OFF/ON
Use arrow keys to cycle these selections.
Note: This option is only available with GSM Modem Option.
TIME & DATE
CLK=01:50:36
DAY=Mon WEEK=2
DATE=21/03/2006
$!4!,/''%2
,/').4%26!,MINS
,/'0%2)/$DAYS
3ET!LARMS
)'./2%&,/7/&&
2%0%!4%VERY-ORNING
)'./2%#!,/&&
2