SECTION 4: TECHNICAL SPECIFICATIONS
Entire Contents Copyright
2016 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
APS CFS300 Series Power Source Operation Manual
Page 31 of 156
PROTECTION MODES
Reverse Current Protection
(RPC)
Displayed if the power source detects negative current feeding
back into the source. RCP will trip when negative power exceeds
75W. Test/Reset LED blinks.
LVP Low Voltage Protection
(LVP)
Displayed if the power source detects a discrepancy between the
output voltage and the set voltage greater than 10V for more
than one second. Test/Reset LED blinks.
4.7
AC Input
AC INPUT SPECIFICATIONS
MODEL
CFS330
CFS360
Input Phases
1ø
1ø or 3ø
Input Voltage
1ø Input
200-240Vac±10%
200-240Vac±10% LN
3ø Input, 3W Delta
200-240Vac±10% LL
3ø Input, 4W Wye
346-416Vac±10% LL
Max. Input Current
23A
1ø: 45A
3ø, 3W: 26A
3ø, 4W: 26A
Max. VA Input Power
4 kVA
8 kVA
Frequency
47 - 63 Hz
Input Power Factor
PFC, > 0.97 @ Full Load
Efficiency
> 78% @ Full Load
AC Input to Output Isolation
Output is galvanically isolated from the power source AC
Input (primary to secondary double insulation)
Isolation Voltage AC Input to Output
3047 Vrms / 4309 Vdc
Note:
AC input current values shown in the table above apply under full power output conditions.
The actual AC input power will be a function of the output power and is shown by model in
the charts on the next page.