CFS300 SERIES OPERATION MANUAL
SECTION 10: CALIBRATION
Entire Contents Copyright
2016 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
APS CFS300 Series Power Source Operation Manual
Page 148 of 156 PN 160952-10
10.3.2
Software Calibration Procedure
To Activate Calibration Mode
To enter the calibration mode power on the unit while holding the 4 key on the numeric
keypad. When in the calibration mode the display will look as follows:
Use the up or down arrow soft keys to navigate to the parameter that you would like to
calibrate. The parameters available for calibration are Voltage 150.0V, Voltage 300.0 V,
Current xx.xxA and A-Peak xx.xA. The actual values for the Current and A-Peak will change
according to the model number.
Use the “
Select”
soft key to select the parameter for calibration. If you press the “
Exit
” soft
key from this screen you will be kicked out of the Calibration mode and returned to the set
screen.
10.3.3
Voltage: 150.0V Calibration
Use the “
∧
,
∨
" soft keys to navigate to the Voltage 150.0V parameter and press the “
Select”
soft key.
Follow the prompt message provided on the display, and press the “
Test/Reset
” button to
move into the calibration screen for voltage. If you press the “
Exit
” soft key at this screen,
you will be returned to the calibration mode screen.