16
Maintain the ViiA
™
7 System
Applied Biosystems ViiA
™
7 Real-Time PCR System Quick Reference Guide
Fast 96-well sample block
• Applied Biosystems ViiA
™
7 System Fast 96-Well Heated Cover (PN 4459838)
• Applied Biosystems ViiA
™
7 System Fast 96-Well Plate Adapter (PN 4459846)
• Applied Biosystems ViiA
™
7 System Fast 96- Well Sample Block (PN 4453559)
Decontaminate the sample block
Clean the sample block to eliminate fluorescent contaminants from the ViiA
™
7 System sample block. Contamination is generally evident
in failed background calibrations where one or more wells consistently exhibit abnormally high signals.
CAUTION!
PHYSICAL INJURY HAZARD
. Do not remove the ViiA
™
7 System cover. There are no instrument components that you
can safely service yourself. If you suspect a problem, contact an Applied Biosystems Service Representative.
CAUTION!
PHYSICAL INJURY HAZARD
. During instrument operation, the sample block can be heated to 100 °C. Before
performing the following procedure, be sure to wait until the sample block reaches room temperature.
CAUTION!
Before using a cleaning or decontamination method other than those recommended by the Applied Biosystems, verify
with Applied Biosystems that the proposed method will not damage the equipment.
Materials required
• Bleach, 10% solution
• Chemwipe
• Cotton or nylon swabs
• Lint-free cloths
• Ethanol, 95% solution
• Safety glasses
• Pipette (100-
μ
L) with pipette tips
• Powder-free gloves
• Screwdriver
• Deionized water
How to handle the sample block
To prevent damaging or contaminating the sample block, handle the assembly as shown. Also, when the
assembly has been removed from the ViiA
™
7 System, place the sample block on a clean, dry surface or
in its shipping container.