Problem
Solution
During the running of the instrument
protocol, the magnet head and the
plastic comb gripper are stuck when
moving down (not moving down
smoothly, but step by step) and there is
an abnormal noise.
Contact the supplier to determine the
problem and replace the coupling if
necessary.
The angles of the turntable plate
positions are misaligned. After plate
position 1 is aligned, the positions of
the other plates are deviated.
Contact the supplier.
After the machine has been running for
long periods of time, the position of
the parts is misaligned. The technician
hits or touches the parts during
operation, causing the misalignment.
The technicians should be careful when
operating and initialize the instrument
before each run.
Breakage of the instrument motor
coupling.
| Temporary solution: Buy new rigid
coupling and replace it after
disassembling the machine.
| Final solution: The new instruments
uniformly use the new rigid coupling to
increase the service life.