The screen displays other pictures
while the program is running, and the
machine is still running. It should return
to normal after a few minutes.
Problem
Solution
The height of the instrument shield
plate is too low, it will hit the plastic
comb on the turntable when the
turntable rotates.
First, check whether it is caused by the
deformation of the shield plate,
manually press or straighten the shield
plate. If it still cannot be solved, you
need to disassemble the machine to
adjust the height of the shield plate
and the position of the coupling.
17
| Temporary solution: For the old
software version, this situation may
occur randomly after the instrument runs
for a long time. Shut down the machine
for a period of time and then restart.
| Final solution: Upgrade the instrument
software to the latest version. No need
to disassemble the machine. Connect
the machine to a computer through the
port on the back of the instrument and
import the latest version of the
software.
Some magnet rods of the magnet head
have broken.
| Temporary solution: Do not add
samples and reagents to the
corresponding well positions of the
broken magnet rods, and continue to use
other well positions to run the program.
| Final solution: Buy a new 96 magnet
head and replace it.