![Apone SiOne Instruction Manual Download Page 19](http://html1.mh-extra.com/html/apone/sione/sione_instruction-manual_2954698019.webp)
Communication Protocol
SiOne Page
18
8: Temperature (Si-S00… and Si-Q… with Si-
S00… versions)
9: Dew point (Si-S00… and Si-Q… with Si-
S00… versions)
10: CO2 (Si-Cx0… and Si-Q… with Si-Cx0…
versions)
11: Differential air pressure (Si-Dx0… versions)
12: Temperature (Si-Dx0 versions)
13: MODBUS – configured via MODBUS
6
DEC
-
7
DEC
Measurement value for minimum output signal -
analog output no. 1 - Si-…A1, Si-…A2, Si-…A3 version
0x03
HEX
0x10
HEX
REAL
-
8
DEC
-
9
DEC
Measurement value for minimum output signal -
analog output no. 2 - Si-…A2, Si-…A3 version
0x03
HEX
0x10
HEX
REAL
-
10
DEC
-
11
DEC
Measurement value for minimum output signal -
analog output no. 3 - Si-…A3 version
0x03
HEX
0x10
HEX
REAL
-
12
DEC
-
13
DEC
Measurement value for maximum output signal -
analog output no. 1 - Si-…A1, Si-…A2, Si-…A3 version
0x03
HEX
0x10
HEX
REAL
-
14
DEC
-
15
DEC
Measurement value for maximum output signal -
analog output no. 2 - Si-…A2, Si-…A3 version
0x03
HEX
0x10
HEX
REAL
-
16
DEC
-
17
DEC
Measurement value for minimum output signal -
analog output no. 3 - Si-…A3 version
0x03
HEX
0x10
HEX
REAL
-
18
DEC
-
19
DEC
Output signal value for sensor error - analog output
no. 1 - Si-…A1, Si-…A2, Si-…A3 version
0x03
HEX
0x10
HEX
REAL
-
20
DEC
-
21
DEC
Output signal value for sensor error - analog output
no. 2 - Si-…A2, Si-…A3 version
0x03
HEX
0x10
HEX
REAL
-
22
DEC
-
23
DEC
Output signal value for sensor error - analog output
Si-…A3 version
0x03
HEX
0x10
HEX
REAL
-
24
DEC
-
49
DEC
Calibration registers (Offset)
0x03
HEX
0x10
HEX
REAL
-
50
DEC
-
75
DEC
Calibration registers (Coeff)
0x03
HEX
0x10
HEX
REAL
-
0
DEC
-
1
DEC
Status register
0x04
HEX
-
DWORD -
2
DEC
-
3
DEC
Humidity (Si-H00… and Si-Q… with Si-H00… versions)
0x04
HEX
-
REAL
-