Rev. A2, 07/16
PG7
17
Automation Products Group, Inc.
Quick Calibration (CALIBR):
Allows the user to perform a quick calibration of the zero and span.
NOTE:
Both zero pressure and full-scale pressure must be applied to the
gauge in order to complete the quick calibration process.
Performing Quick Calibration:
Step 1:
Press the
On/Off
button to enter the main setup menu.
Step 2:
Press
(Z)
to cycle through the menu options until
AdVSET
is
displayed.
Step 3:
Press
(S)
to enter the Advanced Settings menu.
Step 4:
Press
(Z)
to cycle through the options until
CALIBR
is displayed.
Step 5:
Press
(S)
to access the
CALIBR
mode. The gauge will display the
word
ZERO
, prompting the user to perform the zero pressure quick
calibration.
Step 6:
Ensure no pressure is applied to the gauge. Press
(Z)
to zero the
gauge. The gauge will then display the word
SPAN
, prompting the user
to perform the full-scale pressure quick calibration.
Step 7:
Apply full jack pressure to the gauge and press the
(S)
button to
complete the quick calibration and return to advanced setup menu.
Step 9:
To exit the advanced setup menu, press
(Z)
until
EXIT
is displayed
and press
(S)
to exit to the main setup menu.