ORP Measurement 

Press 

 to power on, clean the probe in distilled water and dry it. And place the probe in the 

tested solution and stir gently and let it stand still. Read the measurement after  

 comes up and 

stays. 

Notes 

1) 

Usually there’s no need to calibrate for ORP testing. When used for a long time, users can test 

it in 222 mV ORP standard buffer. If the error is big, please calibrate according to the steps in 

5.1 to 5.4. 

2)  The meter has self-diagnosis function. Please refer to the followed chart for detailed 

information. 

Symbol 

Self-Diagnosis information 

How to fix 

 

 

 

 

 is pressed before the 

measurement becomes stable (smiley 
face comes up) 

Wait  till  the  smile  icon 

 appears  and 

stays, then short press 

 

 

The cleaning and activation of ORP probes 

After long-time use, the surface of the platinum sensor can be contaminated and cause inaccuracy 

and slow response time. If that happened, the followed methods can be adopted to clean and 

activate: 

a)  For contamination caused by inorganic substances, dip the probe into 0.1 mol/L dilute 

hydrochloric acid for 30 minutes. Clean it with distilled water, soak it into the KCL storage 

solution for 6 hours, and then it will be ready for use again. 

b)  For contamination caused by organic substances and oiliness, clean the platinum surface with 

dishwashing liquid, and wash with distilled water. And then the probe will be ready for use again 

after 6 hours’ soaking in the storage solution. 

c)  If the platinum surface is so severely contaminated that there forms an oxide film, toothpaste 

can be used to clean the platinum surface, and then wash it with distilled water. The probe will 

be ready for use again after 6 hours’ soaking in the storage solution. 

Parameter Setting 

9.1  Setting chart 

Symbol 

Parameter Setting Contents 

Code 

Factory Default 

P1 

Low value measurement alarm 

-1000 to 1000 mV 

-1000 

P2 

High value measurement alarm 

-1000 to 1000 mV 

1000 

P3 

Select automatic lock 

Off 

– On 

Off 

P4 

Select backlight 

Off - 1

 

- On 

1 (1 min auto off) 

P5 

Select temperature unit 

˚C - ˚F 

˚C 

P6 

Restore to Factory Default 

No

 

– Yes 

No 

Summary of Contents for Premium ORP5

Page 1: ...ORP5 Premium Pocket ORP Tester APERA INSTRUMENTS Europe GmbH www aperainst de v3 1 ...

Page 2: ... the tester s cap while making sure to push all the way down The tester s waterproof design may be compromised if the cap is not fitted correctly 2 Keypad Functions Short press 2 seconds Long press 2 seconds 1 Short press to turn on the tester and long press to turn off the tester 2 When turned off long press to enter parameter setting 3 In measurement mode short press to turn on backlight 1 In ca...

Page 3: ...ottle 3 Short press to turn the meter on Long press to enter calibration mode If you need to exit calibration mode short press 4 Dip electrode in the standard buffer solution stir gently and allow it to stand still in the calibration solution until a stable reading is reached When stable icon comes up press to adjust the calibration value according to the table below also on the buffer bottle s la...

Page 4: ...d to clean and activate a For contamination caused by inorganic substances dip the probe into 0 1 mol L dilute hydrochloric acid for 30 minutes Clean it with distilled water soak it into the KCL storage solution for 6 hours and then it will be ready for use again b For contamination caused by organic substances and oiliness clean the platinum surface with dishwashing liquid and wash with distilled...

Page 5: ...mV stable icon displays on LCD Red back lit heads up displays B Automatic lock P3 Select On to activate auto lock function When reading is stable for more than 10 seconds the tester will lock the value automatically and HOLD icon will display on LCD Press key to cancel autohold C Backlight P4 Off turn off backlight On turn on backlight 1 backlight will last for 1 minute D Temperature Unit P5 Selec...

Page 6: ...epair or replace free of charge at option of APERA INSTRUMENTS Europe GmbH any malfunctioned or damaged product attributable to responsibility of APERA INSTRUMENTS Europe GmbH for a period of TWO YEARS SIX MONTHS for the probe from the delivery This limited warranty does NOT cover any damages due to Transportation storage improper use failure to follow the product instructions or to perform any pr...

Reviews: