background image

 

Complete kit 

 

Preparation Before Use 

1)  Pour certain amount of Storage solution (KCl) into the probe cap, soak the probe for 3 to 5 

minutes to activate the sensor (if frequently used, there is no need to soak). 

2)  The soaking solution is 3M KCl solution, which comes with the test kit. Users can also make 

it by themselves by dissolving 25g KCl in 100 ml distilled water. 

ORP Calibration 

1)  Clean electrode in distilled water and dry it.     

2)  Pour certain amount of 222 mV ORP standard buffer solution into the calibration bottles 

(about half volume of the bottle). 

3)  Short press  

  to turn the meter on. Long press  

  to enter calibration 

mode. If you need to exit calibration mode, short press 

 

4)  Dip electrode in the standard buffer solution, stir gently, and allow it to stand 

still in the calibration solution until a stable reading is reached. When stable 

icon 

comes up, press 

  to adjust the calibration value according to 

the table below(also on the buffer bottle

’s label).  

 

2

22 mV (25˚C) ORP Standard Buffer Solution Calibration Reference Table 

Temperature (˚C) 

10 

15 

20 

25 

30 

35 

38 

40 

mV 

242 

235 

227 

222 

215 

209 

205 

201 

 

F

or example, if the temperature is around 25˚C, then adjust the calibration value to 222 mV. If the 

temperature is around 20˚C, then adjust the calibration value to 227 mV.  

press 

  to complete the calibration. 

Summary of Contents for Premium ORP5

Page 1: ...ORP5 Premium Pocket ORP Tester APERA INSTRUMENTS Europe GmbH www aperainst de v3 1 ...

Page 2: ... the tester s cap while making sure to push all the way down The tester s waterproof design may be compromised if the cap is not fitted correctly 2 Keypad Functions Short press 2 seconds Long press 2 seconds 1 Short press to turn on the tester and long press to turn off the tester 2 When turned off long press to enter parameter setting 3 In measurement mode short press to turn on backlight 1 In ca...

Page 3: ...ottle 3 Short press to turn the meter on Long press to enter calibration mode If you need to exit calibration mode short press 4 Dip electrode in the standard buffer solution stir gently and allow it to stand still in the calibration solution until a stable reading is reached When stable icon comes up press to adjust the calibration value according to the table below also on the buffer bottle s la...

Page 4: ...d to clean and activate a For contamination caused by inorganic substances dip the probe into 0 1 mol L dilute hydrochloric acid for 30 minutes Clean it with distilled water soak it into the KCL storage solution for 6 hours and then it will be ready for use again b For contamination caused by organic substances and oiliness clean the platinum surface with dishwashing liquid and wash with distilled...

Page 5: ...mV stable icon displays on LCD Red back lit heads up displays B Automatic lock P3 Select On to activate auto lock function When reading is stable for more than 10 seconds the tester will lock the value automatically and HOLD icon will display on LCD Press key to cancel autohold C Backlight P4 Off turn off backlight On turn on backlight 1 backlight will last for 1 minute D Temperature Unit P5 Selec...

Page 6: ...epair or replace free of charge at option of APERA INSTRUMENTS Europe GmbH any malfunctioned or damaged product attributable to responsibility of APERA INSTRUMENTS Europe GmbH for a period of TWO YEARS SIX MONTHS for the probe from the delivery This limited warranty does NOT cover any damages due to Transportation storage improper use failure to follow the product instructions or to perform any pr...

Reviews: