Battery Installation 

Please install batteries according to the following steps. 

*

Please note direction of batteries:  

All POSITIVE SIDES (“+”) FACING UP. (Wrong installation of batteries will 

cause damage to the tester and potential hazards) 

 

 

 Pull the battery cap up 

 Slide the battery cap along to the direction of arrow 

 Open the battery cap 

 Insert the batteries 

(ALL POSITIVE SIDES FACING UP)

 (see graph) 

 Close the battery cap 

 Slide and lock the battery cap along to the direction of arrow 

 

Fit the tester’s cap while making sure to push all the way down. The tester’s waterproof 

design may be compromised if the cap is not fitted correctly. 

 

Keypad Functions 

          Short press 

— < 2 seconds 

          Long press 

— > 2 seconds 

 

 

 

 

 

 

 

1.  Short press to turn on the tester and long press 

to turn off the tester. 

2.  When turned off, long press to enter parameter 

setting. 

3.  In measurement mode, short press to turn on 

backlight. 

 

 

 
 
 

 

1.  In calibration mode, press to change calibration 

value (unidirectional)             

2.  In mode setting, short press to change parameter 

(unidirectional) 

 

1.  Long press to enter calibration mode. 
2.  In calibration mode, short press to con- firm 

calibration. 

3.  When measured value is locked, short press to 

unlock. 

-  - 

-  - 

Summary of Contents for Premium ORP5

Page 1: ...ORP5 Premium Pocket ORP Tester APERA INSTRUMENTS Europe GmbH www aperainst de v3 1 ...

Page 2: ... the tester s cap while making sure to push all the way down The tester s waterproof design may be compromised if the cap is not fitted correctly 2 Keypad Functions Short press 2 seconds Long press 2 seconds 1 Short press to turn on the tester and long press to turn off the tester 2 When turned off long press to enter parameter setting 3 In measurement mode short press to turn on backlight 1 In ca...

Page 3: ...ottle 3 Short press to turn the meter on Long press to enter calibration mode If you need to exit calibration mode short press 4 Dip electrode in the standard buffer solution stir gently and allow it to stand still in the calibration solution until a stable reading is reached When stable icon comes up press to adjust the calibration value according to the table below also on the buffer bottle s la...

Page 4: ...d to clean and activate a For contamination caused by inorganic substances dip the probe into 0 1 mol L dilute hydrochloric acid for 30 minutes Clean it with distilled water soak it into the KCL storage solution for 6 hours and then it will be ready for use again b For contamination caused by organic substances and oiliness clean the platinum surface with dishwashing liquid and wash with distilled...

Page 5: ...mV stable icon displays on LCD Red back lit heads up displays B Automatic lock P3 Select On to activate auto lock function When reading is stable for more than 10 seconds the tester will lock the value automatically and HOLD icon will display on LCD Press key to cancel autohold C Backlight P4 Off turn off backlight On turn on backlight 1 backlight will last for 1 minute D Temperature Unit P5 Selec...

Page 6: ...epair or replace free of charge at option of APERA INSTRUMENTS Europe GmbH any malfunctioned or damaged product attributable to responsibility of APERA INSTRUMENTS Europe GmbH for a period of TWO YEARS SIX MONTHS for the probe from the delivery This limited warranty does NOT cover any damages due to Transportation storage improper use failure to follow the product instructions or to perform any pr...

Reviews: