- 6 -

6. Probe Cleaning

6.1  The tester is only as accurate as the probe is clean. Always thoroughly rinse off the probe before  

and after each measurement with pure water in a container or with a wash bottle. 

6.2 For tough contaminants, detach the sensor shield, soak the probe in Apera probe cleaning  

solution   (AI1166) or detergent water for 30 minutes. Then use a soft brush to remove the 
contaminants. Afterwards, soak the probe in 3M KCL soaking solution for at least 1 hour. Rinse it 
off, then re-calibrate the tester before using again.

7. Probe Storage

7.1  Always store the spear probe in the 3M KCL soaking solution when not in use. Never let it dry out. 

Replace the soaking solution if it gets contaminated.

7.2  NEVER store the probe in pure water like tap, RO, distilled, or deionized water as they could 

damage the pH probe. Pure water is only for rinsing the probe.

5.5  Pure Water pH Measurement
     When testing pure water like tap water, drinking water, RO water or distilled water, it will take 

longer for the readings to get fully stabilized (typically 1-5 minutes). Please be patient. If still 
not working, add Apera 3M KCL (AI1107) to your pure water at the ratio of 1:1000 (e.g. 1 ml KCL 
to 1000 ml water) to accelerate stabilization while minimizing pH change. If the accuracy is not 
meeting your requirement, please contact us at [email protected] to find the specialized meter 
designed for pure water pH test.

8. Parameter Settings

Symbol

Parameter Setting Contents

Code

Factory Default

P1

Select pH buffer solution

USA – NIST

USA

P2

Low value measurement alarm setting

0 to 14.00pH

0

P3

High value measurement alarm setting

0 to 14.00pH

14.00

P4

Select automatic lock

Off – On

Off

P5

Select backlight

Off - 1 - On

1

P6

Select temperature unit

℃ - ℉

P7

Restore to factory default

No – Yes

No

8.1   Setup Menu

Summary of Contents for PH60S

Page 1: ...1 Premium Series PH60S Spear pH Tester pH mV Temperature User Manual IP67 ISO 9001 2015 APERA INSTRUMENTS LLC aperainst com...

Page 2: ...aning 6 7 Probe Storage 6 8 Parameter Settings 6 9 ORP Measurement 7 10 Technical Specifications 8 11 Other Specifications 8 12 Probe Replacement 9 13 Notes 9 14 Battery Replacement 10 15 Troubleshoot...

Page 3: ...hort press to cancel calibration 1 In measurement mode short press to switch between pH and mV 2 In settings short press to change parameters Unidirectional 1 Long press to enter calibration mode 2 In...

Page 4: ...pH 7 00 buffer solution first and make a quick stir in the solution then hold still 4 1 4 Long press to enter calibration mode screen will turn green Short press if you decide to quit calibration and...

Page 5: ...quirement 0 1 pH 2 pt Option A 1st pt 7 00 pH 2nd pt 4 00 pH or 1 68 pH Option A 1st pt 6 86 pH 2nd pt 4 01 pH or 1 68 pH Range 7 00 pH Option B 1st pt 7 00 pH 2nd pt 10 01 pH or 12 45 pH Option B 1st...

Page 6: ...y could damage the pH probe Pure water is only for rinsing the probe 5 5 Pure Water pH Measurement When testing pure water like tap water drinking water RO water or distilled water it will take longer...

Page 7: ...on screen screen turns red 8 3 3 Automatic Lock P4 Select On to activate auto lock function When reading is stable for more than 10 seconds the tester will lock the value automatically and HOLD icon w...

Page 8: ...mp Measuring Range 0 50 C 32 122 F Resolution 0 1 C Accuracy 0 5 C 11 Other Specifications Screen 3 color LCD screen Blue Measurement Green Calibration Red Alarm Reading Lock HOLD Low Voltage Warning...

Page 9: ...PH60S E spear pH probe for soft solid samples AI1203 PH60F E flat pH probe for surface test AI1201 PH60 E regular pH probe for general water solutions AI3711 E PH60 DE double junction pH probe for gen...

Page 10: ...E Battery MUST FACE UP WRONG INSTALLATION OF BATTERIES WILL CAUSE DAMAGE TO THE TESTER AND POTENTIAL HAZARDS Loosen the battery cap lock Pull off the battery cap Slide and unlock battery compartment O...

Page 11: ...t s within the 6 month electrode warranty contact us for warranty fulfillment If there is visible damage replace the electrode Instrument defect Contact us for warranty fulfillment Reading keeps jumpi...

Page 12: ...ons beyond our reasonable control To get the fastest warranty fulfillment go to support aperainst com and click New Support Ticket on the upper right corner Type your email in the requester field Warr...

Reviews: