- 7 -

8.2   Parameter Setup
        When  turned  off,  long  press 

  to  enter  parameter  settings  →  Short  press 

                                                                      

to switch P1-P2-P3…P7 → Short press 

 to select parameter (starts flickering) → Short press 

 to change parameter → Short press 

 to confirm the change → Long press 

 to return to 

measurement mode.

8.3  Parameter Setting Instruction
8.3.1   Select standard pH buffer solution (P1): 
         There are two options of standard buffer solutions: USA series and NIST series. Factory default 

is USA series, for details see section 4.2.

8.3.2  Alarm Function (P2&P3):
     a)  Alarm triggered when measurement ≤ 3.20 pH:
          Preset lowest value (P2) = 3.20 pH, highest value (P3) = 14.00 pH, when measured value ≤ 3.20 

pH (stable   displays on screen); screen turns red.

     b)  Alarm triggered when measurement ≥ 8.60 pH:
        Preset highest value (P3) = 8.60 pH, lowest value (P2) = 0.00 pH, when measured value ≥ 8.60 

pH (stable   displays on screen); screen turns red.

     c)  Alarm triggered when measurement ≤ 3.20 pH or ≥ 8.60 pH
         Preset lowest value (P2) = 3.20 pH, highest value (P3) = 8.60 pH, when measured value is lower 

than 3.20 pH or higher than 8.60 pH (stable   displays on screen), screen turns red.

8.3.3  Automatic Lock (P4): Select “On” to activate auto lock function. When reading is stable for 

more than 10 seconds, the tester will lock the value automatically, and HOLD icon will display on 
LCD. Short press 

 key again to cancel the lock and keep measuring.

8.3.4  Backlight (P5): “Off”-turn off backlight, “On”-always turn on backlight, “1”- backlight will last for 

1 minute.

8.3.5  Temperature Unit (P6): Select between C˚ and F˚.

8.3.6  Factory default setting (P7): Select “Yes” to recover instrument calibration to the theoretical 

value (pH value in zero potential is 7.00pH, slope is 100%), parameter setting return to initial 
value. This function can be used when instrument does not work properly in calibration or 
measurement. Calibrate and measure again after recovering the instrument to factory default 
status.

9. ORP Measurement

ORP stands for Oxidation-Reduction Potential. ORP is a measure of the cleanliness of the water & 
its ability to break down contaminants. An ORP probe is needed to test ORP (sold separately, SKU: 
AI1207). After powering on the tester, press 

 to enter ORP mode (mV). Rinse the probe in distilled 

water and dry it. Dip the probe in your sample solution, make a quick stir, and hold still. Record the 
reading after it is fully stabilized.

Summary of Contents for PH60S

Page 1: ...1 Premium Series PH60S Spear pH Tester pH mV Temperature User Manual IP67 ISO 9001 2015 APERA INSTRUMENTS LLC aperainst com...

Page 2: ...aning 6 7 Probe Storage 6 8 Parameter Settings 6 9 ORP Measurement 7 10 Technical Specifications 8 11 Other Specifications 8 12 Probe Replacement 9 13 Notes 9 14 Battery Replacement 10 15 Troubleshoot...

Page 3: ...hort press to cancel calibration 1 In measurement mode short press to switch between pH and mV 2 In settings short press to change parameters Unidirectional 1 Long press to enter calibration mode 2 In...

Page 4: ...pH 7 00 buffer solution first and make a quick stir in the solution then hold still 4 1 4 Long press to enter calibration mode screen will turn green Short press if you decide to quit calibration and...

Page 5: ...quirement 0 1 pH 2 pt Option A 1st pt 7 00 pH 2nd pt 4 00 pH or 1 68 pH Option A 1st pt 6 86 pH 2nd pt 4 01 pH or 1 68 pH Range 7 00 pH Option B 1st pt 7 00 pH 2nd pt 10 01 pH or 12 45 pH Option B 1st...

Page 6: ...y could damage the pH probe Pure water is only for rinsing the probe 5 5 Pure Water pH Measurement When testing pure water like tap water drinking water RO water or distilled water it will take longer...

Page 7: ...on screen screen turns red 8 3 3 Automatic Lock P4 Select On to activate auto lock function When reading is stable for more than 10 seconds the tester will lock the value automatically and HOLD icon w...

Page 8: ...mp Measuring Range 0 50 C 32 122 F Resolution 0 1 C Accuracy 0 5 C 11 Other Specifications Screen 3 color LCD screen Blue Measurement Green Calibration Red Alarm Reading Lock HOLD Low Voltage Warning...

Page 9: ...PH60S E spear pH probe for soft solid samples AI1203 PH60F E flat pH probe for surface test AI1201 PH60 E regular pH probe for general water solutions AI3711 E PH60 DE double junction pH probe for gen...

Page 10: ...E Battery MUST FACE UP WRONG INSTALLATION OF BATTERIES WILL CAUSE DAMAGE TO THE TESTER AND POTENTIAL HAZARDS Loosen the battery cap lock Pull off the battery cap Slide and unlock battery compartment O...

Page 11: ...t s within the 6 month electrode warranty contact us for warranty fulfillment If there is visible damage replace the electrode Instrument defect Contact us for warranty fulfillment Reading keeps jumpi...

Page 12: ...ons beyond our reasonable control To get the fastest warranty fulfillment go to support aperainst com and click New Support Ticket on the upper right corner Type your email in the requester field Warr...

Reviews: