The Force Dual Phase™ Operation Manual
7
Revision Date: 4/22/2019
2.3.3.
Manual Screen
2.3.3.1. The Manual Screen is used quite often and is where to go to perform any
operations other than normal operation, such as clearing a clogged orifice,
recovering CO
2
before the entered run time is complete, opening the Extractor
Vessel or changing a CO
2
bottle.
Figure 5. Manual Screen: Manual Mode Selected (Reverse Flow)
2.3.3.2. On the Manual Screen, there's a button to switch between automatic flow reversal
and manual flow reversal. When in manual flow reversal mode, the operator can
press a button to switch the flow direction at any time during the run.
2.3.3.3. When in automatic flow reversal, the machine will only change flow direction when
a new extraction is started, meaning that forward flow at the beginning will remain
forward flow until the next run. The next run will be reverse flow and will remain
reverse flow for the entire run.
2.3.3.4. The Manual Screen is used quite often and is where to go to perform any
operations other than normal operation, such as clearing a clogged orifice,
recovering CO
2
before the entered run time is complete, opening the Extractor
Vessel, or changing a CO
2
bottle.
2.3.3.5. The yellow buttons for “reset” of water flow lost, oil pressure lost, VFD fault,
Diaphragm ruptured, or E-Stop will only be visible when the corresponding
abnormal condition exists. If there’s a fault or the system won’t start, user should
check here for reset buttons and check the alarm screen for information about
faults.