OPERATING GUIDELINES 

SOLDERING IRON MANUAL CALIBRATION 

1.  Set soldering iron to desired working temperature. 

2.  Wait  a  few  minutes  for  the  temperature  to  stabilize before  checking 

the temperature difference with an external calibrated probe. 

3.  Access  the  manual  calibration  hole  by  removing  the  rubber  cover  of 

the calibration hole. 

4.  Use a small screw to slowly adjust the trimmer potentiometer thru the 

calibration hole. 

5.  When  tip  temperature  has  been  recalibrated.  Replace  rubber    cover 

into position.

 

SOLDERING IRON DISPLAY GUIDE 

“Pen”— 

Soldering iron is not detected.

 

“OFF”— 

Soldering iron function is currently off.

 

“Hot”— 

Soldering iron function is currently off but the soldering iron 

tip is still very hot.

 

“­ ­ ­ ­ ”— 

Soldering iron is currently in sleep mode.

 

“t00”— 

Soldering iron sleep timer adjustment mode, sleep timer off.

 

“t60”— 

Sleep timer adjustment mode, sleep timer  set at 60 minutes.

 

“001 or –00”— 

Soldering iron digital calibration adjustment mode.

 

“Err”— 

Soldering iron error.

 

SOLDERING IRON SMOKE ABSORBER 

1.  Turn   on   Soldering   Iron   function   and   set   to   desired   working 

temperature. Wait for soldering iron to reach set temperature. 

2.  Turn on the Smoke Absorber function switch.

 

Note: 

Smoke absorber function cannot be enabled if the SMD rework 

function switch is turned “ON”. To utilize the smoke absorber function 

the  SMD rework function switch should be in the “OFF” position. 

7

 

A1 

Air pressure knob

 

A2 

Soldering iron temperature display

 

A3 

Soldering iron temperature knob

 

A4 

Soldering iron calibration hole.

 

A5 

Reset hot air gun temperature

 

A6 

Hot air gun increase temperature adjustment button.

 

A7 

Hot air gun decrease temperature adjustment button.

 

B1 

Soldering iron function switch

 

B2 

Smoke absorber function switch

 

B3 

SMD rework function switch

 

C1 

Air pressure indicator

 

C2 

Set temperature display (Hot air gun)

 

C3 

Actual temperature display (Hot air gun)

 

D1 

Soldering iron terminal

 

D2 

Vacuum outlet

 

D3 

Hot air gun output

 

PANEL CONTROLS

 

A1 

A2 

A3 

A5 

A4 

A6 

B1 

B2 

B3 

C1 

C2  C3 

D1 

D2 

D3 

A7

 

aoyue

   

OFF 

ON 

LOW 

CAL 

HIGH

 

REPAIRING SYSTEM 

TEMPERATURE 

SOLDERING 
IRON 

SOLDERING 

IRON 

SMOKE 

ABSORBER 

SMD 

REWORK 

5

 

VACUUM 

7

Summary of Contents for 968A+

Page 1: ...rsonnel or damage to items in the workplace may result if not carefully used Please read the contents on how to use the equipment and observe the following in order to maximize usage After opening the...

Page 2: ...rflow level is noticeably weaker SOLUTION Try to press the Reset button on the panel and let the device re detect the proper operating frequency Note that resetting the device will also reset all prev...

Page 3: ...ading is not increasing or decreasing based on desired level set temperature The panel will then display a blinking OFF on display panels C2 and C3 afterwards SOLUTION The heating element may be broke...

Page 4: ...s Easily adjust temperature offset with a few simple adjustments Intelligent error reporting mechanism Detects and informs the personnel for problems with the sensor and heating element Auto cooling f...

Page 5: ...nt and reconnect the terminal Be careful not to rub or touch together the heating element wire 6 Reconnect the ground wire after replacing the heating element 7 Re assemble the handle NOTE The life ex...

Page 6: ...te temperature when soldering 2 Cleaning Always clean the soldering tip before using Remove any residual solder or flux that are still adhering to the tip Use a clean and moist cleaning sponge to remo...

Page 7: ...error SOLDERING IRON SMOKE ABSORBER 1 Turn on Soldering Iron function and set to desired working temperature Wait for soldering iron to reach set temperature 2 Turn on the Smoke Absorber function swi...

Page 8: ...ration value into the memory 3 Set soldering iron to desired working temperature 4 Wait a few minutes for the temperature to stabilize before checking the temperature difference with an external calib...

Page 9: ...st the calibration number of the soldering iron Turn SMD and Smoke absorber function switch off Turn Soldering iron function switch on Press and hold the hot air temperature increase button A6 9 SMD R...

Page 10: ...A2 will show the current set temperature based on the knob position 11 OPERATING GUIDELINES 5 After a few seconds the digital display A2 will switch to showing the actual temperature You may start sol...

Reviews: