Manuel d’intructions NTA47248-300A2
96
5.6
X-Scaling setting
The calibrator can measure voltage or current and simply display it or convert it using
the “XScale” function and then display the measured value via a conversion factor.
•
Press the [MENU] key. The following page is shown:
MENU
Co n f i g u r a t i o n
Un i t
Ra mp - Me mo r y s c a n
Da t a l o g g e r
Ad v a n c e d
Pr e s s <
> t o SET
Select the “Unit” option
Un i t
Te mp e r a t u r e
Pr e s s u r e [ mb a r ]
Xs c a l e
Pr e s s <
> t o SET
Select the “XScale” option.
XSc a l e
X1
X2
X3
X4
Pr e s s <
> t o SET
Select one of the 4 available functions. (e.g. X1)
X1
Un i t [ °C]
Pa r a me t e r [ 5 0 mA]
I n L o w [ 4 . 0 0 0 ]
I n Hi g h [ 2 0 . 0 0 0 ]
Di s p L o w [ 0 . 0 ]
Pr e s s <
> t o SET
Select and program each of the 8 variables:
−−−−
Unit: 5 characters length can be programmed for the engineering unit
−−−−
Parameter: select the voltage or the current input range [200mV, 2V, 20V,
50mA]
−−−−
In Low: Change this value with the Zero scale
−−−−
In High: Change this value with the Input Full Scale
−−−−
Disp Low: Change this value with the Displayed low scale
−−−−
Disp High: Change this value with the Displayed High scale
−−−−
Function: Change this value with the linearization function [linear, square,
sqrt, log]
−−−−
Decimals: Change this value with the number of digits to be displayed [0,
1, 2, 3, 4]
•
Press the [ESC] key to return in measuring mode.
5.7
Cycle & Ramp
The output channel can be programmed to simulate a ramp.
The procedure uses the Channel 2 output configuration.
NOTE
BEFORE RUN THE PROCEDURE THE CH2 HAS TO BE SET AS OUTPUT (SEE
PAR. 4.6)
•
Press the [MENU] key. The following page is shown:
Summary of Contents for Calys 120 IS
Page 19: ...Manuel d intructions NTA47248 300A2 19 Voie 2 Simulation Génération ...
Page 60: ...Manuel d intructions NTA47248 300A2 60 12 CERTIFICATS 12 1 Certificat CE 12 2 Certificat ATEX ...
Page 121: ...Manuel d intructions NTA47248 300A2 121 13 CERTIFICATES Declaration of Conformity ...
Page 125: ...Manuel d intructions NTA47248 300A2 125 15 CERTIFICATES ATEX Certificate ...