
Manuel d’intructions NTA47248-300A2
103
Scroll down and select "Alarm"
Co n f i g u r a t i o n
Da t e / t i me
Re j e c t i o n [ 5 0 Hz ]
Ba u d r a t e [ 9 6 0 0 ]
OC Te s t [ En a b l e ]
Al a r m
Pr e s s <
> t o SET
Al a r m
Ge n e r a l [ En a b l e ]
Us e r
Pr e s s <
> t o SET
•
Select General and then Enable or Disable the general system alarms
NOTE: when you disable the alarms, at next turning on the selection will be
lose and alarms become active.
See par.10.1 about error messages.
•
Select User to set alarms definable by the user:
Us e r
On Of f [ o f f ]
Ch a n n e l [ c h 1 ]
Mi n [ 0 ]
Ma x [ 1 0 0 ]
Pr e s s <
> t o SET
•
Set the channel where to have threshold alarms with Channel
•
Set Min and Max threshold
NOTE: unit measure is referred to the used channel
•
Enable the alarm with “OnOff”
NOTE: in case of threshold overcoming, an acoustic signal is emitted until the
signal comes back return under threshold or the alarm is disabled.
5.15
Switching on settings (Power On)
It is possible to set how the instrument turns on through the “Power On” function
•
Press [MENU]
MENU
Co n f i g u r a t i o n
Un i t
Ra mp - Me mo r y s c a n
Da t a l o g g e r
Ad v a n c e d
Pr e s s <
> t o SET
Select “Configuration”
Co n f i g u r a t i o n
Di s p l a y
Po we r On [ Se t Ch ]
Da t e / t i me
Re j e c t i o n [ 5 0 Hz ]
Ba u d r a t e [ 9 6 0 0 ]
Pr e s s <
> t o SET
Select “Power On”
> Po we r On <
Sa f e
Se t Ch
Se t v a l
Pr e s s <
> t o SET
•
Safe: Ch1 = voltage, Ch2 = off. This way avoids any risk of damage for the
instrument
•
Set Ch: it set the last channel settings
•
Set val: it set the last channel settings and the channel 2 output value
Summary of Contents for Calys 120 IS
Page 19: ...Manuel d intructions NTA47248 300A2 19 Voie 2 Simulation Génération ...
Page 60: ...Manuel d intructions NTA47248 300A2 60 12 CERTIFICATS 12 1 Certificat CE 12 2 Certificat ATEX ...
Page 121: ...Manuel d intructions NTA47248 300A2 121 13 CERTIFICATES Declaration of Conformity ...
Page 125: ...Manuel d intructions NTA47248 300A2 125 15 CERTIFICATES ATEX Certificate ...