For veterinary use only
-
19 -
Confirm that the operation mode can be switched by pressing the button and checking the operation mode
indicator.
Press the SWITCH button as described in the Section 4.4.15, “Switch operation mode between flexible and
rigid endoscope mode”; confirm that the operation mode indicator turns red when rigid endoscope mode is
selected;.
Press the SWITCH button again to switch back to flexible endoscope mode, and the indicator turns green
accordingly.
The light source incorporates an air pump and water from the water bottle to feed air and water into the body
cavity from the nozzle at the distal end of endoscope. Confirm that air and water is fed from the nozzle at the
endoscope’s distal end.
Press the airflow switch to turn ON the airflow feeding.
Immerse the distal end of the insertion section in the sterile water; confirm that no air bubbles are emitted
when the air/water valve is not operated.
Cover the hole of the air/water valve of the endoscope to feed air; confirm that air bubbles continuously
emitted from the air/water nozzle.
Uncover the hole of the air/water valve of the endoscope
Remove the distal end of the insertion section from the sterile water.
Depress the air/water valve of the endoscope; confirm that water exiting from the air/water nozzle.
Release the air/water valve of the endoscope.
Press the power switch of the endoscope imaging processor.
Confirm that the endoscope imaging processor is turned OFF and the power indicator dims.
3.7 Inspection of the operation mode switch (only for VET-OR1200R)
3.8 Inspection of the air and water feeding
3.9 Inspection of power OFF
⚫
If the power indicator remains illuminated after turning the endoscope imaging processor OFF,
do not use it, unplug the power cord from the wall mains outlet and contact AOHUA.
1
2
1
2
1
2
3
4
5
6
7