- 25 -
F6
(
Relay
)
Menu
Parameter Choices
Descritpion
Level 2 Submenu
0 (NTEP)
1 (OMIL)
2 (CANADA)
F6.1
(
mode
)
3 (NONE)
For OIML, NTEP and CANADA application, Tare
removal is only allowed when Gross = 0. When
NONE is selected, Tare removal can be done at any
weighing mode.
For NTEP and OIML, a new Tare can be acquired
even when there is a stored Tare.
For CANADA, a new Tare can be acquired after the
stored Tare is removed.
For NONE, NTEP and CANADA, when the current
weight is within the specified zero range, zero scale
can be performed irrespective it is in Gross or Net
mode.
For OIML, zero scale can only be performed when it
is in Gross mode and Tare can be removed when it
is in Net mode.
F6.2
(
P1
)
Number
Specifies fixed value 1. Refer to Section 6 for Relay
Output setup.
F6.3
(
P2
)
Number
Specifies fixed value 2. Refer to Section 6 for Relay
Output setup
3.2.7 F7(Ver) Menu
Table 3-7 F7(Ver) Menu
Fig. 3-10 F7(Ver) Menu