NOTES:
Press the
B5
soft key from the Option Sub-Menu
(page 2-18) to toggle between feet and meters. Val-
ues entered in either will freely convert to the other.
Loss and relative propagation velocity values for
many common cable types are listed in the table on
the inside of the manual back cover.
Making Power Measurements
Power measurement is accomplished using a broadband (1 MHz to
3 GHz) RF detector, P/N 5400-71N50. The power monitor displays the
measured power in dBm or Watts.
Entering Power Monitor Mode
Step 1.
Press the POWER soft key, from the Main Menu
(page 2-8)
Zeroing the Power Monitor
Step 2.
With no power applied to the DUT, press the ZERO
soft key from the Power menu (page 2-19). Wait a
few seconds while the Site Master accumulates sam-
ples of the quiescent power level. When complete,
ZERO ADJ: ON is displayed in the message area.
2 - 40
Chapter 2 Operation