then subtracted from subsequent measurements
prior to display.
q
MAIN
— Returns to the Main Menu.
Remote Operation
All Site Master functions, settings, and operating modes (except for
power on/off) are controllable using commands sent from an external
computer via the serial port. A detailed description of the programming
commands is available in the Site Master Programming Manual, P/N
10580-00012.
Operating Procedures
Site Master operation can be divided into four functional areas: com-
mon functions, frequency domain, distance domain, and power moni-
tor.
•
Common functions are options and scale functions that are
common to both frequency- and distance-domain measure-
ments.
•
Frequency domain measurements consist of Return Loss (RL),
Standing Wave Ratio (SWR), and Cable Loss (CL) made over
a selectable frequency range. RL and SWR characterize the
magnitude of reflections present in a device or transmission
line. CL measurements display the insertion loss of a transmis-
sion line over frequency.
•
Distance domain measurements—commonly known as
distance-to-fault (DTF)—are made over a selectable distance
range. They include RL or SWR, but they also return informa-
2 - 20
Chapter 2 Operation