33
7.1
AIM OF THE MACHINE
The machine has been designed to carry out temperature and humidity tests on components and electrical equipment.
The machine is also suitable for temperature and humidity tests on industrial and pharmaceutical products (CHAL-
LENGE -E series).
Before placing any object whatsoever inside the chamber, make sure that the object itself will not be damaged by the
test temperature.
For further information, please refer to chapter 3 entitled “Technical specifications”.
7.2
MACHINE END-USE
The machine has been designed to be used in companies and must only treat the objects mentioned in paragraph 7.1.
Before treating different materials from those indicated in paragraph 7.1, please contact the manufacturer to obtain a
specific authorization.
7.3
OPERATOR
No special technical knowledge is needed to use this machine.
The operator should read this handbook carefully. However, experience with the use of electrical and electronic
equipment would be preferable.
Special attention must be paid to the temperatures which can be reached inside the test chamber. The
temperature range for the equipment given in this handbook ranges from +180
°
C to-20
°
C or -40
°
C or -70
°
C.
Before opening the door to handle the tests, set the equipment to values close to ambient temperature and
give enough time for temperature to stabilize itself; wear special gloves that resist at high/low temperatures.
7
USE FORESEEN BY THE MANUFACTURER
7.5
USE OF PROTECTIVE CLOTHING
Suitable protective clothing should be worn according to norms in force in order to protect:
- against the temperatures of the test chamber and of the test materials if time is not given for the temperature to reach
acceptable values.
7.6
DANGEROUS AREAS AND WASTE RISKS
During tests personnel should not enter the chamber or touch the test products; the chamber holds risks at
both high and low temperatures.
Take all the necessary precautions and wear suitable protective clothing.
If the chamber is opened at high or low temperatures, the gaskets could be damaged.
CH0305-202
CH0305-204
7.4
USE LIMITS
The machine can only be used by specialized personnel.
The machine can be switched on only once the compressor oil has been heated.
If the machine has been left unused for over 48 hours, the oil heating procedure
must be carried out as described in paragraph 8.1.
The machine may be used within the temperature and humidity range described
in chapter 3 entitled “TECHNICAL SPECIFICATIONS”.
This series of machines must not contain:
-
inflammable and/or explosive liquid or solid materials.
-
liquid or solid materials which could generate inflammable gases during
evaporation or sublimation.
-
liquid or solid materials containing chemically aggressive substances.
-
liquid or solid materials could generate aggressive fumes during evapora-
tion or sublimation.
Always set the relevant software and hardware alarm devices
before carrying out a temperature cycle (paragraph 8.2).