CH0305-260
6.12
5.4
RS232 configuration
RS232/485 IN
RS232/485 OUT
1
2
3
4
5
6
7
8
9
○
○
○
○
remote control (PC)
CH0305-282
RS 232/485 IN connector (5.4) (fig. 6.12)
Serial interface
This connector is usually used for the remote control of the
chamber by means of a PC with WINKRATOS
®
software
by Angelantoni S.p.A.
RS 232/485 OUT connector (5.5) (fig. 6.13)
This connector is used only with the RS485 configuration
to connect several machines to each other and to use a
single remote control with one PC (fig. 6.16).
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
CH0305-272
6.14
5.4
5.4
5.4
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
6.16
Only with
RS485
configuration
PC
WINKRATOS SYSTEM
STANDARD
CONFIGURATION
The WINKRATOS system is able to operate several ma-
chines at the same time.
Each RS232 computer serial port (PC) is connected to each
RS232 connector (paragr. 5.4 fig. 6.14).
As shown in the fig. 6.16 the connector can also be used in the
standard RS 485 for a MULTIDROP connection of several
machines on-line.
IN THIS CASE THE CONFIGURATION OF THE
MACHINE’S INTERNAL MICROCONTROLLER MUST
BE SET UP BY THE MANUFACTURER.
The RS 485 standard can also be used with
special software that could also be realized by
the user.
The WINKRATOS system does not operate this
standard.
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
CH0305-260
6.13
5.5
RS485 configuration
RS232/485 IN
RS232/485 OUT
○
○
○
○
remote contro (PC)
CH0305-283
CH0305-270
5.4
5.5
5.4
5.5
PC
CH0305-273
PC WITH USER’S
SOFTWARE
RS232
M
6.15
I f the user needs to manage the chamber in his own way, he can project his own dedicated software and connect
a computer to the connector 5.4 (fig. 6.16). To cope with this need it is necessary to expressly require to the
manufacturer the documentation relative to the
communications protocol.
The communication protocol allows to know the data exchanging mode between the user’s software and the
machine M (fig. 6.15).
CONNECTION WITH THE USER’S SOFTWARE
PC WITH
WINKRATOS
5.4
MULTIDROP SYSTEM-
THE WINKRATOS SYSTEM
DOES NOT OPERATE THIS STANDARD
1
2
3
4
5
6
7
8
9
TX
RX
TX-/RX-
TX+/RX+