8
seconds starts.
Step 8
Your blood glucose result will appear on the screen after 5 seconds. The result
will be automatically saved in meter simultaneously.
Note:
To ensure accurate results, make sure the confirmation window of the
Test Strip is completely filled with your blood sample.
Step 9
The Glucose AG-605A results range between 20 mg/dL to 600 mg/dL(1.1
mmol/L to 33.3 mmol/L). If a “HI” message appears on your display, your
Glucose Meter has detected that your blood glucose level is higher than
600mg/dL or 33.3mmol/L. It is suggested that you review your testing
procedure and test again with a new Glucose Test Strip to confirm the result. If
the same result occurs, consult your healthcare professional immediately.
Step 10
If a “LO” message appears on your display, your Glucose Meter has detected
that your blood glucose level is lower than 20mg/dL or 1.1mmol/L. It is
suggested that you review your testing procedure and test again with a new
Glucose Test Strip to confirm the result. If the same result occurs, consult your
healthcare professional immediately.
After approximate 2 minutes, if no further tests are performed, the Glucose
Meter will automatically turn off.
Step 11
Remove the test strip from the meter and discard it. It is strongly suggested
that you use a tissue to remove the used test strip. Discard the used lancet
properly in accordance with local
、
state
、
and federal law. After remove the test
strip from the test port, the result goes on showing for one minute and then the