background image

 

meter  will  turn  off  automatically  or  you  can  long  press  the  S  button  after 
remove the used test strip to turn the meter off  immediately or insert another 
test strip to begin a new test. 

NOTICE: 

To  remove  the  Test  Strip  from  the  Glucose  Meter  for  proper  disposal

We 

recommend using a tissue paper to remove the Test Strip form the Glucose Meter. 

 

Note:  According  to  the  clinical  research,  people  without  diabetes  are 
those whose blood glucose level fallen within 70mg/dL

110 mg/dL. 

 
Control Solution Test 

Material needed: 

  Glucose Meter 

  Test strip 

  Control solution 

If  long  press  the  S  button  for  about  two seconds in  step5,  the  CTL will 
appear  on  the  screen  which  means  that  the  test  is  under  the  CTL 
condition and the result will not be stored this time. 

     

 

Note: 1. Do not use the expired control solution. 

            2.  The  control  solution  are  not  necessary  accessory  for  customers,  please 

contact us if you need.

 

 

Recall Test Results 

1.  With the meter turned off, long press and release M button.           
2.  The first test result with date and time will automatically appear. 

     

 

   

 

3.  Short  press  M  button  once  and  next  most  recent  test  result  will  appear. 

Continue short pressing M button to recall the last 350 test results in order. 
“End” will be displayed after the last result showed and the meter will power 

Summary of Contents for AG-605A

Page 1: ...m if it is not working properly or if it has suffered any damage Do not let the Blood Glucose Monitoring System come into contact with surfaces which are too hot to touch Do not block test ports nor p...

Page 2: ...mm L 50mm W 22mm H 3 Weight 60g exclude batteries 4 Measuring method Amperometric technology using glucose oxidase 5 Measuring range 20 mg dL 600 mg dL 1 1 mmol L 33 3mmol L 6 Measurement unit mg dL o...

Page 3: ...test strip Avoid making the liquid into the test strip and the button Avoid getting water or any other solution inside the meter Blood Sample The fresh capillary vessel whole blood finger tip blood or...

Page 4: ...ize batteries and confirms well positive and negative extremely 3 Slide battery cover back into place Note Discard the used battery according to the local law Test Strip The Test Strips are designed t...

Page 5: ...the date setting Short press and release the MEM button until the correct date appears 5 With the correct date on the display short press and release the S button to bring up the time setting The hou...

Page 6: ...ired Testing Your Blood Glucose Level Before testing you will need the following items Glucose AG 605A Meter AGS 1000 test strip Adjustable Automatic Lancing Device Sterile Lancet CAUTION To reduce th...

Page 7: ...Replace the end cap back on the body of the lancing device 5 Pull back on the cocking mechanism until it comes to a stop 6 Place the lancing device firmly against the sample site and press the release...

Page 8: ...f the same result occurs consult your healthcare professional immediately Step 10 If a LO message appears on your display your Glucose Meter has detected that your blood glucose level is lower than 20...

Page 9: ...t strip Control solution If long press the S button for about two seconds in step5 the CTL will appear on the screen which means that the test is under the CTL condition and the result will not be sto...

Page 10: ...aning the andon Blood Glucose Monitoring System gently wipe the exterior surface using a damp soft cloth AVOID USING ANY ORGANIC SOLVENT for cleaning Discard the used lancet properly in accordance wit...

Page 11: ...countdown The meter is calculating the result At the end of the count down the meter will display the test result No action required Blood glucose level is lower than 20mg dl 1 1mmol L The message in...

Page 12: ...12 Display the result Do nothing Demonstrates the memory the serial number Do nothing Memory result demonstration Do nothing No memory in meter Do nothing Result display is end Do nothing...

Page 13: ...libration Please contact us Use a used test strip Please use a new strip and re test Meter internal memory unit error Reinsert the CODE bottle please contact us if it appear for three time Environment...

Page 14: ...e number printed on the Test Strip vial Incorrect Code numbers has been inputted into the Meter Insert the new CODE card to match the Code number on the Test Strip vial label Test results are inconsis...

Page 15: ...15 15minutes before performing a new test The Meter countdown did not start Test Strip has not been inserted correctly Use a new Test Strip and redo the test...

Page 16: ...erature limitation Expired Date Anti garbage Lotus Global Co Ltd 47 Spenlow House Bermondsey London SE16 4SJ Andon Health Co Ltd Register Address No 3 JinPing Street Ya An Road Nankai District Tianjin...

Reviews: